Operating and Programming
Introduction
Introduction
This section describes the additions, changes, and deletions to the information that appears in the introduction of the 5071A Operating and Programming manual.
Introducing the 5071A
5071A Major Features
Option 048: In the bulleted text delete:
•45 minute standby battery
Operating The 5071A
Clock Functions
Any TCO Option: 1st paragraph, 2nd sentence is changed:
The 1 pps output is moveable in time and can be automatically synchronized with other 1 PPS sources.
Any TCO Option: 3rd paragraph is changed:
1 pps output: There is one 1 pps output on the 5071A front panel.
Any TCO Option: 4th paragraph is changed:
The 1 pps output is TTL compatible and requires 50Ω termination. You can automatically synchronize the 1 PPS signal to an external 1 PPS, or manually change its timing by up to ± 500 ms in 50 ns steps. Use the
Option 048: 2nd paragraph (page vii) is changed:
Power input and management: The 5071A operates from either ac or dc power.
1.External dc power: when below the specified minimum voltage (40 Vdc on both dc inputs), or when absent, the 5071A switches to,
2.Ac line power.
(Item 3 is deleted.)