MODEL 381pH/ORP

SECTION 3.0

 

MAINTENANCE

3.3 MODEL 381

3.3.1Platinum Electrode Check. The platinum elec- trode may be checked as follows: There are two types of standard solutions which may be used to check the ORP electrode/transmitter system.

Type 1: One type of commonly used ORP standard solution is the saturated quinhydrone solution. A com- monly used ORP standard solution is a saturated quin- hydrone solution. This can be made by simply adding a few quinhydrone crystals to either a 4 pH or a 7 pH buffer. Quinhydrone is only sightly soluble so only a few crystals will be required. The solution will have a yellow color. The resulting potentials should be within ±20 millivolts of the value shown in Table 3-1. The ORP value of saturated quinhydrone solution is not stable for long periods of time and therefore new solutions should be made each time.

CAUTION

The solution used during the following check is an acid and should be handled with care. Follow the directions of the acid manufacturer. Wear the proper protective equipment. If con- tact with skin of clothing is made, immediately rinse with plenty of clean water.

Type 2: A second ORP standard solution is the Ferric- Ferrous Ammonium Sulfate Solution (PN R508-16OZ), and it can be ordered as a spare part; otherwise, it can be prepared from the following recipe: Dissolve 39.2 grams of reagent grade ferrous ammonium sulfate, Fe(NH4)2 (SO4)2 • 6H2O and 48.2 grams of reagent grade ferric ammonium sulfate, FeNH4(SO4)2 • 12H2O, in approximately 700 milliliters of water (dis- tilled water is preferred, but tap water is acceptable).

Slowly and carefully add 56.2 milliliters of concentrated sulfuric acid. Add sufficient water to bring the total solu- tion volume up to 1000 ml. This standard ORP solution, although not as simple to prepare as the quinhydrone recipe, is much more stable, and will maintain its milli- volt value for approximately one year when stored in glass containers. This solution (ferric/ferrous ammoni- um sulfate) will produce a nominal ORP of 476 +20 mV at 25°C when used with a saturated KCl/AgCl refer- ence electrode and platinum measuring electrode. Some tolerance in mV values is to be expected due to the rather large liquid reference junction potentials which can arise when measuring this strongly acidic and concentrated solution. However, if the measuring electrodes are kept clean and in good operating condi- tion, consistently repeatable calibrations can be carried out using this standard solution.

3.3.2Cleaning Platinum Electrode. The electrode is never exposed to these undesirable compounds. In the event poisoning is suspected, the electrode can be restored to normal operation by simply cleaning the plat- inum electrode with baking soda. Polish it by rubbing it with a damp paper towel and baking soda until a bright, shiny appearance is attained.

TABLE 3-1. ORP of Saturated Quinhydrone Solutions

 

 

 

 

pH 4

 

 

 

pH 7

 

 

 

 

 

 

 

 

 

 

 

Temperature °C

 

20

25

30

 

20

25

30

Millivolt Potential

 

268

264

260

 

94

87

80

 

 

16

Image 20
Contents PH/ORP Sensor Essential Instructions Model 381 pH/ORP Sensor About This Document Section Description and Specifications Features and ApplicationsSensor Specifications Materials of ConstructionMeasuring Range pH Pressure Rating 790 kPa abs at 100C 100 psig at 212FExample Page Section Installation Model 381 Sensor Component Locator Diagram Section Submersion Installation Diagram Flow Through Installation Diagram Code 03 or Retro-Fit Kit Code 04 Flow Powered Cleaner, Model 40038105 40038103 40038116 Calibration START-UP and CalibrationModel 381 ORP Section Maintenance Temperature C Millivolt Potential 268 264 260 Troubleshooting Probable Cause Remedy Element Temperature vs Resistance of AutoTemperature Accessories Model 381 Spare PartsPage Section Ultrasonic Cleaner Operation 40038156 40037505 NON Warranty Repair Section Return of MaterialReturn of Materials Request Warranty Return of MaterialSpecifications subject to change without notice