MODEL 381pH/ORP | SECTION 3.0 |
| MAINTENANCE |
TABLE 3-1. Troubleshooting
| Trouble | Probable Cause Remedy |
| ||
|
|
|
|
|
|
Meter reads off scale. | Reference electrolyte depleted. | Replace reference electrolyte as |
| ||
|
| instructed in Section 3.2.2. |
| ||
|
|
|
|
| |
| Defective preamplifier. | Check preamplifier as instructed in |
| ||
|
| Section 3.4 and replace |
| ||
|
| preamplifier if defective. |
| ||
|
|
|
|
| |
| T.C. element shorted. | Check T.C. element as instructed in |
| ||
|
| Section 3.5, and replace body if |
| ||
|
| defective. |
| ||
|
|
|
|
| |
| Electrode not in solution or sample | Make sure sensor is in solution or that |
| ||
| stream is not full. | sample stream is full (refer to Section |
| ||
|
| 2.0 for installation details). |
| ||
|
|
|
|
|
|
| Open glass electrode. | Replace glass electrode. |
| ||
|
|
|
|
| |
| Plugged liquid junction. | Replace liquid junction. |
| ||
|
| Recharge reference cell. |
| ||
|
|
|
|
| |
| Reference element open- | Replace reference element. |
| ||
| no contact. |
|
|
|
|
|
|
|
|
|
|
Meter reads between 3 and 6 pH | Electrode cracked. | Replace electrode. |
| ||
regardless of actual pH of |
|
|
|
|
|
solution or sample. |
|
|
|
|
|
|
|
|
|
|
|
Meter indication swings widely | T.C. element shorted. | Check T.C. element as instructed in |
| ||
in AUTO T.C. Mode. |
| Section 3.5 and replace body if |
| ||
|
| defective. |
| ||
|
|
|
|
| |
Span between buffers extremely | T.C. element open. | Check T.C. element as instructed in |
| ||
short in AUTO T.C. mode. |
| Section 3.5 and replace body if |
| ||
|
| defective. |
| ||
|
|
|
|
| |
Sluggish or slow meter indication | Electrode coated. | Clean electrode as instructed in |
| ||
for real changes in pH level. |
| Section 3.2.1. |
| ||
|
|
|
|
| |
| Electrode defective. | Replace electrode. |
| ||
|
|
|
|
| |
Transmitter cannot be | Electrode coated or cracked. | Clean electrode as instructed in |
| ||
standardized. |
| Section 3.2.1 and, if cracked, |
| ||
|
| replace electrode. |
| ||
|
|
|
|
| |
| Reference chamber is contaminated | Replace gel solution as |
| ||
|
| instructed in Section 3.2.2. |
| ||
|
|
|
|
| |
| Defective preamplifier. | Check preamplifier as instructed in |
| ||
|
| Section 3.4 and replace if |
| ||
|
| defective. |
| ||
|
|
|
|
| |
| If trouble still persists, | Replace |
| ||
| electrode is defective. |
|
|
|
|
|
|
|
|
| |
Meter short spans. | Old glass electrode (greater than 1 | Replace electrode. |
| ||
| year use). |
|
|
|
|
|
|
|
|
| |
| Coated glass. | Clean with soft cloth and clean water. |
| ||
|
|
|
|
|
|
17