Emerson 381 instruction manual Retro-Fit Kit Code 04 Flow Powered Cleaner, Model

Page 14

MODEL 381pH/ORP

SECTION 2.0

 

INSTALLATION

PRESSURE DROP AND RECOMMENDED FLOW

RATES- FLOW POWERED CLEANER (CODE 04)

kPa PSIG

24020 POSSIBLE ELECTRODE DAMAGE

220

ABOVE THE THIS POINT

PRESSURE DROP

200 15

RECOMMEND-

180 ED OPERATING

10 RANGER

160

140

5

120INEFFECTIVE CLEANING BELOW THIS POINT

100

 

 

 

 

 

 

 

 

gpm @ 60°

1

2

3

4

5

6

7

8

 

 

 

 

 

 

 

 

 

 

 

 

 

5 10 15 20 25 30 liters/min

FLOW RATE

FIGURE 2-6. Retro-Fit Kit Code 04 Flow Powered Cleaner, Model 381

10

Image 14
Contents PH/ORP Sensor Essential Instructions Model 381 pH/ORP Sensor About This Document Section Description and Specifications Features and ApplicationsMeasuring Range pH Sensor SpecificationsMaterials of Construction Pressure Rating 790 kPa abs at 100C 100 psig at 212FExample Page Section Installation Model 381 Sensor Component Locator Diagram Section Submersion Installation Diagram Flow Through Installation Diagram Code 03 or Retro-Fit Kit Code 04 Flow Powered Cleaner, Model 40038105 40038103 40038116 Calibration START-UP and CalibrationModel 381 ORP Section Maintenance Temperature C Millivolt Potential 268 264 260 Troubleshooting Probable Cause Remedy Element Temperature vs Resistance of AutoTemperature Accessories Model 381 Spare PartsPage Section Ultrasonic Cleaner Operation 40038156 40037505 NON Warranty Repair Section Return of MaterialReturn of Materials Request Warranty Return of MaterialSpecifications subject to change without notice