Sierra Monitor Corporation 4105 manual Sensor Type Units, Range Resolution Response Time

Page 14

Instruction Manual

8.SPECIFICATIONS (Continued)

Model

4105-12

4105-21

4105-22 4105-253

4105-263

Gas

NO2

HCL

HCN

NH

HF

Sensor Type

E.C.

E.C.

E.C.

3

E.C.

E.C.

Units

ppm

ppm

ppm

ppm

ppm

Range

0-20

0-20

0-20

0-50

0-10

Resolution

0.2

0.5

0.1

0.1

0.5

Response Time

35 sec.

150 sec

70 sec.

45 sec.

30 sec.

to 90% of signal

 

 

 

 

 

Sensor Life 2

2 yrs

2 yrs

2 yrs

N/A

N/A

Operating Range

 

 

 

 

 

Temperature

14 to 122oF

14 to 122oF

14 to 122oF 14 to 113oF

14 to 113oF

 

-10 to 50oC

-10 to 50oC

-10 to 50oC

-10 to 45oC

-10 to 45oC

Relative Humidity

15-90%

15-90%

15-90%

20-90%

20-90%

Pressure

10%

10%

10%

10%

10%

Electrical Data

 

 

 

 

 

Input Voltage DC

 

 

19-30 VDC at less than 100 mA

Output-Normal

 

 

4-20 mA DC linear

Output-Trouble

 

 

 

0 mA

 

Construction

 

 

 

 

 

Dimensions

 

H: 7.6", D: 4.6", W: 4.2" (19.3 x 11.7 x 10.7 cm)

Weight (Module)

 

 

 

5.0 lb. (2.2 kg)

Mounting

 

 

 

3/4" NPT

 

Housing

 

 

Explosion proof

(NEMA 7) / NEMA 4X Enclosure (Div. I, Class 1, Groups B, C, D)

Notes: 2. Sensor life is for use at standard temperature and pressure with occasional exposure to the gas of interest 3. Diffusion via membrane. Dimensions are H: 10.2", D: 6.0", W: 6.0" (25.9 x 15.2x 15.2 cm)

Model 4105 Gas Sensor Module (12/02)

Page: 11

Image 14
Contents Model 4105 Series Applicability & Effectivity Table of Contents Section Product Description Recommended Wire/Cable Gauge Introduction WiringWire Gauge Maximum Length Preventative MaintenanceQuick Start Operation Transmitter Cover Plate Model 4105 Gas Sensor Module 12/02Calibration Service 4a. Combustibles 4b LEL Sensor Fault Supervision4d Missing Electromechanical Sensor Fault Supervision 4c Electrochemical Oxygen and ToxicsPower Supply InstallationExplosion Proof Installation Sensor LocationsToxics Oxygen Wiring Diagram Model 4105 Gas Sensor Module 12/02Combustible Specifications Range Resolution Response Time Sensor Type UnitsHCN Replacement Parts LimitedwarrantyAppendix C End of Sensor Life Indication Appendix a Unity Gain ModeAppendix B L.C.D. Readout Calibration Procedure Appendix C Combustible Gas Scaling Factors Table Combustible GAS Scaling Factors Factor