Sierra Monitor Corporation 4105 manual Product Description

Page 4

1.PRODUCT DESCRIPTION

1.1 Introduction

Model 4105 Series Gas Sensor module features non- intrusive calibration, 4-20 mA output signal and local continuous display of the gas concentration in ppm (toxics), % LEL (Combustibles), or % vol (Oxygen). It is designed for use with industry standard control in- struments.

This manual provides instructions for 4105 series gas sensor modules that utilize common packaging and transmitter electronics with different sensors for de- tection of various gases. The full model number of the gas sensor module includes a suffix, 4105-XX, where “XX” is a number that identifies a gas type.

1.2 Application

The Model 4105-XX Gas sensor is intended for use in ambient monitoring applications. It is designed for fixed installation and for continuous operation.

Optional fittings and adapters can be supplied by Si- erra Monitor to provide continuous sample delivery to the sensor module.

1.3 Configuration

The gas monitor is comprised of a NEMA-7/4 enclo- sure which contains the transmitter electronics and a gas sensor which is installed in one of the two 3/4" conduit hubs.

5.42"

 

 

 

4.50"

 

 

2.81" 1.75"

 

 

FAULT /KEYPAD

 

 

MODEL 4105-02

 

 

 

Combustibles % LEL

 

5.50"

Unity SPAN

ZERO

 

Ph. 408-262-6611

 

Milpitas, CA

7.25"

 

CAL

Figure 1.2 Outline View - Gas Sensor Module

Instruction Manual

Figure 1.1 Model 4105 Gas Sensor Module

1.3.1 Electronics

When installed, the transmitter electronics will be con- nected to a power supply and control device via three wire cable.

The electronics provide a 4-20 mA current loop which is proportional to the full sensitivity of the sensor. Integral features include:

-L.C.D. readout of ppm of gas concentration.

-Magnetic switches for non-intrusive calibration.

-Visual indicators for fault condition.

-Electrical fault (0 mA) to controller.

The transmitter electronics assembly includes a printed circuit assembly and a cover plate which con- tains the L.C.D. readout and magnetic switches for calibration.

1.3.2 Sensor

The gas sensor cell is enclosed in a threaded housing and is plugged into a signal card located in the enclo- sure. This configuration allows for easy field replace- ment of the sensor cell.

Combustible: The gas sensor is a catalytic bead type.

Toxics & Oxygen: The gas sensor is an Electro Chemi- cal type.

Model 4105 Gas Sensor Module (12/02)

Page: 1

Image 4
Contents Model 4105 Series Applicability & Effectivity Table of Contents Section Product Description Introduction Wiring Wire Gauge Maximum LengthRecommended Wire/Cable Gauge Preventative MaintenanceQuick Start Operation Transmitter Cover Plate Model 4105 Gas Sensor Module 12/02Calibration Service 4b LEL Sensor Fault Supervision 4d Missing Electromechanical Sensor Fault Supervision4a. Combustibles 4c Electrochemical Oxygen and ToxicsInstallation Explosion Proof InstallationPower Supply Sensor LocationsCombustible Oxygen Wiring Diagram Model 4105 Gas Sensor Module 12/02Toxics Specifications Range Resolution Response Time Sensor Type UnitsHCN Replacement Parts LimitedwarrantyAppendix B L.C.D. Readout Calibration Procedure Appendix a Unity Gain ModeAppendix C End of Sensor Life Indication Appendix C Combustible Gas Scaling Factors Table Combustible GAS Scaling Factors Factor