
5.CALIBRATION
5.1 Frequency of Calibration
The manufacturer recommends that the gas sensor module be calibrated every ninety days.
5.2 Calibration Process
The output signal of the gas sensor module is cali- brated using a span mixture containing a known con- centration of the gas of interest. The concentration of the span gas must be within the full scale of the sensor module and should be either 50% of the full scale, or approximately equal to the lowest alarm level.
Calibration requires application of the span gas to the sensor and adjustment of the “SPAN” magnetic switch making the module signal output and display equiva- lent to the concentration of sample gas. The
5.3 Equipment Required
The following tools and equipment will be required for calibration:
-Magnetic tool
-Gas Sensor Calibrator, Model
-Calibration Gas
-Disbursing Calibration Adapter (Model 5358- 01)
5.4Calibration Procedure
Routine calibrations are easily performed using the magnetic tool provided with each sensor.
Briefly hold the magnet tool close to the small dot located on the lower edge of the front panel. The arrow on the upper left side of the LED will illuminate and the
Instruction Manual
Connect the calibration adapter and expose the sen- sor to an appropriate SPAN gas using a span mixture containing a known concentration of the gas of inter- est at a minimum flow rate of 300 cc/minute. (Use per- meation tube for
If the L.C.D. does not display the correct SPAN value, a SPAN adjustment is required. With the arrow still flashing, hold the magnet close to the UP SPAN or
The monitor is now calibrated. Deactivate calibration by holding the magnet close to the small dot again. This releases the 1.5 mA lock.
5.5 Fault Identification
A Fault condition is detected if the sensor output drifts far enough negative to cause the
Model | Gas | Min. Flow Rate |
|
| (cc/min) |
CH4 | 100 | |
O2 | 100 | |
CO | 150 | |
H2S | 300 | |
Cl2 | 300 | |
H2 | 300 | |
SO2 | 300 | |
NO2 | 300 | |
HCl | 300 | |
HCN* | 300 | |
NH3 | 400 | |
HF | 500 |
*Note: SO2 may be used instead of HCN to
calibrate: 10ppm SO2 = 16 ppm HCN
*Note: H2S may be used instead of HCl to
calibrate: 10 ppm H2S = 20 ppm HCl
** Note: Use permeation tube for
Table 4.1
Calibration Span Gas Flow Rates
Model 4105 Gas Sensor Module (12/02) | Page: 5 |