Omega LV800 manual Troubleshooting, I N T E N a N C E

Page 8

T R O U B L E S H O O T I N G

PROBLEM: Sensor will not sense material

CAUSE/SOLUTION:

1)Verify power is applied to the sensor.

2)Verify "Sense Adjust." Reposition adjustment clockwise therefore making the probe more sensitive to "difficult to sense" materials.

3)Verify probe coverage when sensing is expected. The sen- sor is not designed to be "tip sensitive." Permit significant probe coverage before expecting material sensing.

4)Verify electrical continuity between sensor housing and ves- sel wall.

5)Verify connection of the electronic module to the probe.

PROBLEM: Sensor remains in the "SENSE" mode even when material is absent

CAUSE/SOLUTION:

1)Verify the active probe is not in direct contact with any inter- nal vessel structure. If so, reposition sensor.

2)Verify "Sense Adjust" setting. Reposition adjustment counter-clockwise therefore making the probe less sensitive to "easy to sense" materials.

3)Verify there is not a large buildup of material between the active probe and the probe housing. See "Cleaning" in the "Maintenance" section of this bulletin.

PROBLEM: Sensor does not seem to hold its calibration

CAUSE/SOLUTION:

1)Verify that the "sense adjust" is positioned at ideal point as described in 2-Step Calibration procedure. Placing setting too close to threshold can create nuisance false signals.

2)Maximize probe coverage of material being sensed. "Difficult to detect" materials (i.e. low dielectric or light mate- rials) are more easily detected when more probe is covered. Avoid setups where "probe tip sensitivity" is required.

3)Insure electrical continuity between the probe's mounting point (3/4" NPT or 1 1-4" NPT) and the bin wall. Sealant tapes (i.e. Teflon tape) can electrically isolate these regions and make sensing more difficult.

4)Assess the temperature swing that the electronics sees in application. Applications with "difficult to detect" materials and that experience a wide temperature swing are the most vulnerable. Consult factory for support.

PROBLEM: Output contacts perform opposite of designations (N/O, N/C)

CAUSE/SOLUTION:

1)Designations on PCB relate to relay status when in "Fail- Safe Low" mode and when no material is sensed. If "Fail- safe High" mode is used, the designations are reversed.

Swap wire terminations of N/O and N/C if necessary. Changing the fail-safe selection is not recommended.

M A I N T E N A N C E

Fuse Replacement:

The fuse incorporated into the LV800 PCB is not intended for operator replacement. A qualified technician can replace the applicable fuse according to the following specifications. If nec- essary, consult the factory for additional technical assistance or for return of the LV800.

115VAC

F1: 5x20mm, 0.63A Time Lag (Slo-Blo), 250VAC

Littelfuse: #218.063

Bussman: #GDC-63mA 230VAC

F1: 5x20mm, 0.32A Time Lag (Slo-Blo), 250VAC

Littelfuse: #218.032

Bussman: #GDC-32mA

Preventive Maintenance:

The LV800 design is virtually maintenance free. In typical applications, once the sensor is properly calibrated, it will oper- ate without any further attention. In abrasive applications, probe condition should be periodically observed. Although the probe's construction is made with high quality engineering plas- tics, heavy abrasion can wear away the probe's insulators. This can influence the calibration and eventually lead to sensor fail- ure. The electronics are housed in a weatherproof enclosure. In addition the PCBs are conformal coated with a silicone- based material to further prevent electrical influence by con- densation. The sensor electronics should periodically be observed for any signs of contamination caused by improper enclosure protection (i.e. insure cover is fully engaged, and that wire entries are properly sealed.)

Cleaning Requirements:

Environmental and hazardous location protection is highly dependent on the proper installation of the cover to the hous- ing. A lubricant is applied to the housing/cover threads to enhance ability to twist-on twist-off cover. Clean and re-lubri- cate threads if necessary to insure trouble-free operation. Do not physically alter threads in anyway otherwise environmental or hazardous location protection could be compromised.

The driven shield feature of the LV800 is designed to minimize the need for cleaning the probe portion of the sensor. In appli- cations where material build-up is so excessive that electronics can not adequately compensate, probe cleaning may be nec- essary. If so, wipe the probe with a clean damp cloth from the housing to the probe tip.

7

Image 8
Contents User’s Guide Servicing North America LV 8 0 0 S E R I E S Cover Removal See Figure E I N S TA L L AT I O N C O N S I D E R AT I O N SChoosing a Location See Figure E C T R I C a L I N S TA L L AT I O N Power Input See L I B R AT I O NOutput Relay Contacts See Figure Step Calibration- probe exposed to materialDelay Troubleshooting I N T E N a N C EC H a N I C a L S F E T Y E C I F I C AT I O N SWARRANTY/DISCLAIMER Shop online at omega.com