6. Specifications

General

 

Flow range:

 

5061 and 5062:

0.11 to 2.6 lpm (0.03 to 0.7 gpm)

5063 and 5064:

1.13 to 12.11 lpm (0.3 to 3.2 gpm)

Linearity:

±1.2% of full range

Repeatability:

±0.5% of full range

Connections:

¼ in. NPT (male) or ISO 7/1 - R¼ (male)

Cable length:

8 m (25 ft.), can splice up to 300 m (1000 ft.) max.

Cable type:

2-conductor twisted pair w/shield (Belden 8451)

Weight:

250 g (9.6 oz.)

Wetted Materials

Sensor body and cover: 40% glass fi lled Polyphenylene Sulfi de

Rotor: PEEK, natural, unfi lled

Rotor O-ring: Viton®

Cover O-ring: Viton®

Electrical

Supply voltage: 5 to 24 VDC, regulated

Supply current:

10 mA max.

Output type:

Open-collector NPN transistor, 10 mA max sink

Quality Standard

• Manufactured under ISO 9001 and ISO 14001

Fluid Conditions

Max. pressure/temperature:

0 to 80°C (32 to 176°F) @ 5.5 bar (80 psi) max

psi bar

251.7

201.4

151.0

100.7

5 0.3

0 0

Pressure Drop Across Sensor vs. Flow Rate

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

FP-5061/FP-5062

 

 

FP-5063/FP-5064

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

1.9

3.8

5.7

7.6

9.5

11.4

L/m

0

0.5

1

1.5

2

2.5

3

gpm

Dimensions

Top View

100 mm (3.94 in.)

Side View

36 mm (1.42 in.)

81 mm

(3.2 in.)

Mounting tabs for metric M3 or standard #6 screws on 68 mm (2.68 in.) bolt circle

OMEGA FP-5060 Series Micro Flow Sensor

page 5

Image 5
Contents User’sGuide Servicing North America Safety Instructions Instrument WiringSeries Sensor Recommended Sensor PositionSurface Mount Instructions Fitting InstallationPipe Pipe fitting FP-5060 Sealant CalibrationSpecifications Omega FP-5060 Series Micro Flow Sensor WARRANTY/DISCLAIMER OMEGA…Of Course