Rite Life 393, 394 manual Exposure Setting

Page 19

activated. When the green color is selected, switches A,B,C,D bracketed as green are activated.

Exposure Setting

A

B

C

D

1

MIN

OFF

OFF

OFF

OFF

2

 

ON

OFF

OFF

OFF

3

 

OFF

ON

OFF

OFF

4

 

OFF

OFF

ON

OFF

5

 

ON

ON

ON

OFF

6

 

OFF

OFF

OFF

ON

7

MAX

OFF

OFF

ON

ON

Adjusting Exposure Setting (394)

The sensitometer has the capability of seven different exposure times for both single and dual. The factory preset exposure setting is “3” for both single and dual. If different exposure times are required, refer to the chart below—or bottom of unit—for the settings. Each exposure setting will move the speed index on the film one step.

The dipswitch used to change the exposure time is located on the bottom of the unit. When single emulsion is selected, switches A,B,C,D bracketed as “SINGLE” are activated. When the dual emulsion is selected, switches A,B,C,D bracketed as “DUAL” are activated.

Exposure Setting

A

B

C

D

1

MIN

OFF

OFF

OFF

OFF

2

 

ON

OFF

OFF

OFF

3

 

OFF

ON

OFF

OFF

4

 

OFF

OFF

ON

OFF

5

 

ON

ON

ON

OFF

6

 

OFF

OFF

OFF

ON

7

MAX

OFF

OFF

ON

ON

13

Image 19 Contents
393/394 Sensitometer Page FCC Proprietary Notice CE DeclarationIii Limited WarrantyTable Of Contents Unpacking and Inspection Getting StartedInstrument Description Applying Power Sensitometer Monitoring For Process Control Application and ProceduresSelection and Use of Film Log E CurveScheduling Sensitometric Control Processing ProcedureManual Data Recording Procedure Daily Plotting of Data on Process Control Record Developer Type Developer vendor identificationPrinting Process Control Records Automatic Data Recording Procedure-Using DensitometerSpeed Index Base+Fog Possible Cause Processor TroubleshootingSetting Exposure Color Using the SensitometerSelecting Single or Dual Exposure 394 only Adjusting Exposure SettingExposure Setting Exposing Film Important Notes Repair Information General MaintenanceCalibration/Recertification General Cleaning Cleaning the InstrumentCleaning the Step Tablet Replacing the Battery Weight Technical SpecificationsPage Page 393-500 Rev. O-4/28/04
Related manuals
Manual 28 pages 15.3 Kb