Rite Life 393, 394 manual Applying Power

Page 9

Applying Power

The sensitometer is shipped from the factory with the battery installed and power switch in the “Off” (left) position—slide switch to the “On” (right) position. The battery switch for the 393 is located on the front of the instrument. The battery switch for the 394 is located on the bottom of the instrument.

Because there is no drain on the battery during non-use times, the power switch can remain in the “On” position. The only time that the power switch must be turned “Off” is when the instrument is transported or the battery replaced.

Your sensitometer is designed to operate from a “9 Volt” alkaline battery for approximately one year. When the battery is low, the circuit will not allow any exposures to occur. Should your unit become inoperative, replace the battery first, refer to Section Five. If this does not rectify the problem, refer the instrument to X-Rite or an authorized service center for proper servicing. There are no user serviceable components in the sensitometer.

3

Image 9 Contents
393/394 Sensitometer Page FCC Proprietary Notice CE DeclarationIii Limited WarrantyTable Of Contents Unpacking and Inspection Getting StartedInstrument Description Applying Power Sensitometer Monitoring For Process Control Application and ProceduresSelection and Use of Film Log E CurveScheduling Sensitometric Control Processing ProcedureManual Data Recording Procedure Daily Plotting of Data on Process Control Record Developer Type Developer vendor identificationPrinting Process Control Records Automatic Data Recording Procedure-Using DensitometerSpeed Index Base+Fog Possible Cause Processor TroubleshootingSetting Exposure Color Using the SensitometerSelecting Single or Dual Exposure 394 only Adjusting Exposure SettingExposure Setting Exposing Film Important Notes General Maintenance Repair InformationCalibration/Recertification Cleaning the Instrument General CleaningCleaning the Step Tablet Replacing the Battery Weight Technical SpecificationsPage Page 393-500 Rev. O-4/28/04
Related manuals
Manual 28 pages 15.3 Kb