Replacing the Battery

A low battery condition prevents exposures from occurring. Always Replace The Battery before referring the unit for service.

1.Turn Power Switch “Off” and remove battery access cover by sliding outward.

2.Disconnect old 9-Volt battery from the circuit and discard.

WARNING: Connecting the battery leads backward may cause circuit failure.

3.Connect a replacement 9 Volt Alkaline battery to the circuit—recognizing proper polarity—and insert into the compartment.

4.Reinstall battery access cover and turn power “On.”

Battery Access Cover

9-Volt Alkaline

Battery

18

Image 24 Contents
393/394 Sensitometer Page FCC CE Declaration Proprietary NoticeLimited Warranty IiiTable Of Contents Getting Started Unpacking and InspectionInstrument Description Applying Power Application and Procedures Sensitometer Monitoring For Process ControlLog E Curve Selection and Use of FilmProcessing Procedure Scheduling Sensitometric ControlManual Data Recording Procedure Developer Type Developer vendor identification Daily Plotting of Data on Process Control RecordAutomatic Data Recording Procedure-Using Densitometer Printing Process Control RecordsProcessor Troubleshooting Speed Index Base+Fog Possible CauseUsing the Sensitometer Setting Exposure ColorAdjusting Exposure Setting Selecting Single or Dual Exposure 394 onlyExposure Setting Exposing Film Important Notes General Maintenance Repair InformationCalibration/Recertification Cleaning the Instrument General CleaningCleaning the Step Tablet Replacing the Battery Technical Specifications WeightPage Page 393-500 Rev. O-4/28/04