Thermo Products 9678BNWP manual Electrode Maintenance

Page 16

Electrode Maintenance

1.Inspect the electrode for scratches, cracks, salt crystal buildup, or membrane/junction deposits.

2.Rinse off any salt buildup with distilled water. Remove any membrane/junction deposits as directed in the General Cleaning section.

3.If the electrode is refillable, drain the reference chamber, flush it with fresh filling solution and refill the chamber with fresh filling solution.

4.If the electrode is refillable, make sure that the filling solution does not crystallize or dry out. Remove any crystals by rinsing the electrode with distilled water or wiping with a damp, lint-free tissue. To prevent crystallization of the filling solution, cover the fill hole whenever the electrode is being stored and open the fill hole when calibrating and measuring.

14

Redox/ORP Electrode User Guide

Image 16
Contents User Guide Page Introduction Epoxy Refillable ORP/ATC Triode Cat. No Bnmd and 9180BNRequired Equipment Sample Requirements Electrode Preparation for Refillable Electrodes Electrode PreparationElectrode Preparation for Gel-Filled Electrodes Refillable ORP Electrode Electrode Calibration ORP Standard ValuesRedox/ORP Electrode User Guide Measurement Techniques Electrode Measurements Direct MeasurementsOxidation-Reduction Titrations References Typical Redox Titration Methods Unknown Interferences Suitable Titrants Titration Reaction Electrode Maintenance General Cleaning Electrode Cleaning ProceduresCleaning Solutions Sure-Flow Electrode Cleaning Cat. No Bnwp Disassembling the Sure-Flow ElectrodeElectrode Storage Short-term Storage up to one dayLong-term Storage over one day Reassembling the Sure-Flow ElectrodeTroubleshooting MeterElectrode Sample/ApplicationAssistance Warranty Checking Electrode OperationOrdering Information Thermo Fisher Scientific