Thermo Products 9678BNWP manual Suitable Titrants Titration Reaction

Page 15

Suitable Titrants

Titration Reaction

Ref

potassium permanganate

5Sb+3 + 2MnO4¯ + 16H+ –›

2

(KMnO4)

5Sb+3 + 2Mn+2 + 8H2O

 

sodium bromate (NaBrO3)

BrO3¯ + 3As+3 + 6H+ –›

3

 

Br¯ + 3As+5 + 3H2O

 

potassium permanganate

5H2O2 + 2MnO4¯ + 6H+ –›

5

(KMnO4)

5O2 + 2Mn+2 + 8H2O

 

sodium iodide (NaI)

in acid solution Cl2 +2I¯ –› I2 +

1

 

2Cl¯, In basic solution

 

 

2ClO¯ + 2I¯ –› I2 + 2Cl¯ + 2OH¯

 

sodium iodate ( NaIO3)

2I2 + IO3¯ + 5HCN + H+ –›

4

 

5ICN + 3H2O

 

sodium thiosulfate

I2 + 2S2O3 -2–› 2I¯ + S4O6 -2

3

(Na2S2O3)

 

 

 

 

 

ceric sulfate [Ce(SO4)2]

Fe+2 + Ce+4 –› Ce+3 + Fe+3

2

(not recommended for

 

 

samples containing Cl¯)

 

 

 

 

 

potassium dichromate

Cr2O7 -2+ 6Fe+2 + 14H+ –›

3

(K2Cr3O7), recommended

2Cr+3 + 6Fe+3 + 7H2O

 

for samples containing Cl¯

 

 

 

 

 

ferrous sulfate (FeSO4)

MnO4¯ + 5Fe+2 + 8H+ –›

2

 

Mn+2 + 5Fe+3 + 4H2O

 

potassium permanganate

3Mn+2 + 2MnO4¯ + 7H2O –›

1

(KMnO4)

5MnO3 + H2O + 4H+

 

ceric sulfate [Ce(SO4)2]

Sn+2 + 2Ce+4 –› Sn+4 + 2Ce+3

4

iodine (I2)

SO2 + I2 + 2H2O –›

1

 

SO4-2+ 2I¯ + 4H+

 

sodium bromate (NaBrO3)

BrO3¯ + 3Ti+ 6H+ –›

3

 

Br¯ + Ti+3 + 3H2O

 

Iodine (I2)

I2 + 2S2O3 -2–› 2I¯ + S4O6

3

 

 

 

potassium dichromate

3UO+2 + Cr2O7 -2+ 8H+ –›

3

(K2Cr2O7)

3UO2 +2 + 2Cr+3 + 4H2O

 

sodium iodate (NaIO3)

ZnHg(SCN)4 + 6IO3¯ + 6Cl¯ +

2

 

8H+ –› Zn+2 + Hg+2 + 6ICl +

 

4HCN + 4SO4 -2

Redox/ORP Electrode User Guide

13

Image 15
Contents User Guide Page Epoxy Refillable ORP/ATC Triode Cat. No Bnmd and 9180BN IntroductionRequired Equipment Sample Requirements Electrode Preparation Electrode Preparation for Refillable ElectrodesElectrode Preparation for Gel-Filled Electrodes Refillable ORP Electrode ORP Standard Values Electrode CalibrationRedox/ORP Electrode User Guide Measurement Techniques Direct Measurements Electrode MeasurementsOxidation-Reduction Titrations References Typical Redox Titration Methods Unknown Interferences Suitable Titrants Titration Reaction Electrode Maintenance Electrode Cleaning Procedures General CleaningCleaning Solutions Disassembling the Sure-Flow Electrode Sure-Flow Electrode Cleaning Cat. No BnwpReassembling the Sure-Flow Electrode Electrode StorageShort-term Storage up to one day Long-term Storage over one daySample/Application TroubleshootingMeter ElectrodeChecking Electrode Operation Assistance WarrantyOrdering Information Thermo Fisher Scientific