Thermo Products 9678BNWP manual Electrode Preparation for Refillable Electrodes

Page 6

Electrode Preparation

Electrode Preparation for Refillable Electrodes

1.Remove the protective shipping cap from the sensing element and save the cap for storage.

2.Clean any salt deposits from the exterior of the electrode by rinsing with distilled water.

3.Uncover the filling hole by removing the tape and then add electrode filling solution, Cat. No. 900011, to the electrode. To maintain an adequate flow rate, the level of filling solution must always be above the reference junction and at least one inch above the sample level. The filling hole should be open whenever the electrode is in use.

4.For Sure-Flow electrode only, wet the junction by pressing down on the electrode cap to let a few drops of the filling solution flow out of the electrode and then replenish any lost solution.

5.Shake the electrode downward (similar to a clinical thermometer) to remove air bubbles.

6.Connect the electrode to the meter.

Note: The fill hole must be open when the electrode is in use.

Electrode Preparation for Gel-Filled Electrodes

1.Remove the protective shipping cap from the sensing element and save the cap for storage.

2.Clean any salt deposits from the exterior of the electrode by rinsing with distilled water.

3.Shake the electrode downward (similar to a clinical thermometer) to remove air bubbles.

4.Connect the electrode to the meter.



Redox/ORP Electrode User Guide

Image 6
Contents User Guide Page Introduction Epoxy Refillable ORP/ATC Triode Cat. No Bnmd and 9180BNRequired Equipment Sample Requirements Electrode Preparation Electrode Preparation for Refillable ElectrodesElectrode Preparation for Gel-Filled Electrodes Refillable ORP Electrode Electrode Calibration ORP Standard ValuesRedox/ORP Electrode User Guide Measurement Techniques Electrode Measurements Direct MeasurementsOxidation-Reduction Titrations References Typical Redox Titration Methods Unknown Interferences Suitable Titrants Titration Reaction Electrode Maintenance Electrode Cleaning Procedures General CleaningCleaning Solutions Sure-Flow Electrode Cleaning Cat. No Bnwp Disassembling the Sure-Flow ElectrodeLong-term Storage over one day Electrode StorageShort-term Storage up to one day Reassembling the Sure-Flow ElectrodeElectrode TroubleshootingMeter Sample/ApplicationAssistance Warranty Checking Electrode OperationOrdering Information Thermo Fisher Scientific