Thermo Products 9678BNWP manual Electrode Storage, Short-term Storage up to one day

Page 19

Reassembling the Sure-Flow Electrode

1.Moisten the O-ring on the electrode body with a drop of filling solution. Insert the screw-thread end of the electrode body into the tapered, ground end of the sleeve.

2.Push the body into the sleeve using a gentle twisting motion until the bottom surface of the inner cone is flush with the tapered end of the sleeve.

3.Place the spring onto the electrode body and screw on the cap. Refill the electrode with filling solution.

Electrode Storage

Short-term Storage (up to one day)

To ensure a quick response and free-flowing junction, the reference junction must not dry out. It is recommended that the electrode be soaked in a 4 M potassium chloride (KCl) solution. Electrodes may also be stored in water.

Long-term Storage (over one day)

For gel-filled electrodes– rinse the electrode with distilled water, dry the electrode with a lint-free tissue and cover the measuring surface with the protective cap.

For refillable electrodes– rinse the electrode with distilled water, drain all of the filling solution from the electrode, flush out the electrode with distilled water, cover the measuring surface with the protective cap and store the electrode dry.

Redox/ORP Electrode User Guide

17

Image 19
Contents User Guide Page Epoxy Refillable ORP/ATC Triode Cat. No Bnmd and 9180BN IntroductionRequired Equipment Sample Requirements Electrode Preparation for Refillable Electrodes Electrode PreparationElectrode Preparation for Gel-Filled Electrodes Refillable ORP Electrode ORP Standard Values Electrode CalibrationRedox/ORP Electrode User Guide Measurement Techniques Direct Measurements Electrode MeasurementsOxidation-Reduction Titrations References Typical Redox Titration Methods Unknown Interferences Suitable Titrants Titration Reaction Electrode Maintenance General Cleaning Electrode Cleaning ProceduresCleaning Solutions Disassembling the Sure-Flow Electrode Sure-Flow Electrode Cleaning Cat. No BnwpReassembling the Sure-Flow Electrode Electrode StorageShort-term Storage up to one day Long-term Storage over one daySample/Application TroubleshootingMeter ElectrodeChecking Electrode Operation Assistance WarrantyOrdering Information Thermo Fisher Scientific