Agilent Technologies 08495-90025 service manual Installation, Initial Inspection Mating Connectors

Page 16

Initial Inspection

NOTE

Mating Connectors

CAUTION

Installation

Installation

Inspect the shipping container for damage. If the shipping container or cushioning material is damaged, it should be kept until the contents of the shipment have been checked for completeness and the instrument has been checked mechanically and electrically. A procedure for checking electrical performance is given under “Operator’s Check” on page 12 (also see “Performance Tests” on page 14). If the contents of the shipment are incomplete, if there is mechanical damage or defect, or if the instrument does not pass the electrical performance test, notify the nearest Agilent Technologies office. If the shipping container is damaged, or the cushioning material shows signs of stress, notify the carrier as well as the Agilent office. Keep the shipping materials for the carrier's inspection. The office will arrange for repair or replacement without waiting for claim settlement.

Containers and materials identical to those used in factory packaging are available through Agilent Technologies offices. If the instrument is being returned to Agilent for servicing, attach a tag indicating the type of service required, return address, model number, and full serial number. Also, mark the container FRAGILE to assure careful handling. In any correspondence, refer to the instrument by model number and full serial number.

Mating RF connectors used with the Option 001 must be type-N male connectors, which comply with U.S. military standard MIL-C-39012 (See Figure 7 on page 19). For Option 002, male SMA connectors must be used. For Option 003, 7 mm mating connectors must be used (Figure 6 on page 18).

The solenoid drive cable connector plug is a Viking Industries, Inc., part number TKP12-102-P-TS-100-AU (see Replaceable parts). The connector plug and contacts without the cable can be ordered from a Viking distributor. When ordering the complete cable assembly from Agilent, use the part number listed in Table 5, “Replaceable Parts,” on page 15.

When installing the instrument, make sure that the connectors do not support weight or bear torque. The preferred procedure is to set up all equipment in position before connecting the instrument.

108494/95/96G/H Operating And Service Manual

Image 16
Contents Agilent Technologies 8494/95/96G/H Attenuators Restricted Rights Legend What You’ll Find in This Manual… Warranty Certification Documentation WarrantyService and Support Safety and Regulatory InformationSafety Earth Ground Before Applying Power Instrument MarkingsSpecification OverviewInstruments Covered by ManualTypical Four Section Attenuator Schematic Diagram 8494G/H Attenuator Sections 8495G/H Attenuator Sections8496G/H Attenuator Switching Order 8496G/H Attenuator SectionsOption Connector Description Instrument OptionsFrequency Range Attenuation SpecificationsAttenuation Accuracy RF Power Handling Maximum SWRMaximum Residual Attenuation CapabilitySolenoid Cable Connector SectionAtten CardPhysical Characteristics CharacteristicInstrument Dimensions Weight Depth x width x height Operating ValueInstallation Initial Inspection Mating ConnectorsInstallation Instructions Operator’s Check Operating InstructionsDescription Meter Indication dB Minimum SWR Meter RangeMaximum Adjustments Performance TestsReplaceable Parts Description Part NumberTroubleshooting ServiceRepair To disconnect To care for the deviceReplacing Amphenol 7 mm Center Contact Dimensions Replacing Connector Outer Shell