Service

3.Allow the jaws of the tool used to close and pull straight back from the connector without twisting. The contact should come with the tool. If not, repeat the process. Do not reuse the contact.

4.Snap in a new contact by pushing it in place. Test the action of the new contact by pushing it in. It should spring out again when released.

Type N Connector For critical dimensions for the type-N connector, see below.

Dimensions

 

Figure 7 Type-N Connector

 

Replacing the

 

Connector Outer Shell

 

 

 

 

NOTE

 

The connector outer shell can be replaced only on the Option 001 (type-N

 

 

female) or the Option 003 (7 mm). The outer shell on the Option 002 (SMA)

 

 

cannot be replaced in the field. If these connectors are damaged, return the

 

 

instrument to Agilent Technologies for repair.

 

 

The connector outer shells on the 0ption 001 and 003 may be replaced as

 

 

 

 

follows:

 

1.

With a 9/16-inch (1/2-inch for 7 mm) thin open-end wrench, unscrew the

 

 

 

outer connector body.

 

2.

Replace the connector outer shell. See Table 5 on page 15 for

 

 

 

replaceable parts numbers.

 

3.

Tighten the connector with the same wrench called out in step 1.

8494/95/96G/H Operating And Service Manual 19

Image 25
Contents Agilent Technologies 8494/95/96G/H Attenuators Restricted Rights Legend What You’ll Find in This Manual… Certification Documentation Warranty WarrantySafety and Regulatory Information Service and SupportInstrument Markings Safety Earth Ground Before Applying PowerOverview SpecificationInstruments Covered by ManualTypical Four Section Attenuator Schematic Diagram 8495G/H Attenuator Sections 8494G/H Attenuator Sections8496G/H Attenuator Sections 8496G/H Attenuator Switching OrderInstrument Options Option Connector DescriptionFrequency Range Attenuation SpecificationsAttenuation Accuracy Maximum SWR RF Power HandlingMaximum Residual Attenuation CapabilitySection Solenoid Cable ConnectorAtten CardCharacteristic Physical CharacteristicsInstrument Dimensions Weight Depth x width x height Operating ValueInitial Inspection Mating Connectors InstallationInstallation Instructions Operator’s Check Operating InstructionsDescription Meter Indication dB Minimum SWR Meter RangeMaximum Performance Tests AdjustmentsDescription Part Number Replaceable PartsTroubleshooting ServiceRepair To care for the device To disconnectReplacing Amphenol 7 mm Center Contact Replacing Connector Outer Shell Dimensions