Agilent Technologies 08495-90025 service manual To disconnect, To care for the device

Page 23

Service

4.Close the gap between coupling nut with the nut on the extended-sleeve connector.

To disconnect:

1. Loosen the coupling nut of the connector showing the wider gold band.

CAUTION

Part the connectors carefully to prevent striking the inner conductor contact.

To care for the device:

1.Keep contacting surfaces smooth and clean. Irregularities and foreign particles can degrade electrical performance.

2.Protect the contacting surfaces when the connector is not in use by leaving the coupling sleeve extended.

3.Use lintless material and/or firm-bristles brush such as a tooth brush for cleaning. If a cleaning fluid is needed, use isopropyl alcohol.

8494/95/96G/H Operating And Service Manual 17

Image 23
Contents Agilent Technologies 8494/95/96G/H Attenuators Restricted Rights Legend What You’ll Find in This Manual… Certification Documentation Warranty WarrantySafety and Regulatory Information Service and SupportInstrument Markings Safety Earth Ground Before Applying PowerManual SpecificationOverview Instruments Covered byTypical Four Section Attenuator Schematic Diagram 8495G/H Attenuator Sections 8494G/H Attenuator Sections8496G/H Attenuator Sections 8496G/H Attenuator Switching OrderInstrument Options Option Connector DescriptionAttenuation Accuracy SpecificationsFrequency Range Attenuation Capability RF Power HandlingMaximum SWR Maximum Residual AttenuationCard Solenoid Cable ConnectorSection AttenOperating Value Physical CharacteristicsCharacteristic Instrument Dimensions Weight Depth x width x heightInitial Inspection Mating Connectors InstallationInstallation Instructions Description Operating InstructionsOperator’s Check Maximum SWR Meter RangeMeter Indication dB Minimum Performance Tests AdjustmentsDescription Part Number Replaceable PartsRepair ServiceTroubleshooting To care for the device To disconnectReplacing Amphenol 7 mm Center Contact Replacing Connector Outer Shell Dimensions