Hach 900 manual Unpacking the Instrument, Selecting the Installation Site

Page 13

Section 3

Installation

DANGER

Only qualified personnel should conduct the installation tasks described in this section of the manual

3.1 Unpacking the Instrument

Remove the sampler from the shipping carton and inspect it for any damage. Contact Hach Customer Service at 1-800-227-4224 if any items are missing or damaged.

3.2 Selecting the Installation Site

The sampler is completely self-contained and may be located indoors or outdoors without protection, between 0 to 50 °C (32 to 120 °F). Follow the guidelines below and Figure 3 to allow complete drainage of the intake line and prevent cross-contamination between samples.

Install the sampler as close to the source as site conditions permit. This will increase pump tube life and optimize overall sampler performance.

Mount the sampler to a stable flat surface using the wall mounting bracket, Cat. No. 3596 (Figure 4).

Install the sampler above the sample source, with the intake tubing sloping downward to the sample.

Make sure the tubing is free of kinks or loops.

Figure 3 Placement of the Sampler

1

2

Flow

1.Slope tubing down to source (no loops, kinks, or excess tubing).

2.Locate strainer in an area of turbulent and well-mixed flow.

 

Page 13

8544install.fm

Installation

Image 13
Contents Sigma 900 Composite Sampler Page Table of Contents Table of Contents Pump/Controller Housing Controller DimensionsSample Volume Sample Volume RepeatabilityPage Instrument Description Interface ConnectorsReceptacle Caps General InformationSection Humidity IndicatorFront Panel Keypad and DisplaySafety Information Safety InformationUse of Hazard Information Precautionary LabelsHazardous Locations Peligro Confined Space EntryDefinition of Confined Space Installation Selecting the Installation SiteUnpacking the Instrument Attaching the Intake Line Mounting the SamplerVinyl Intake Tubing Teflon-Lined Polyethylene Intake LineInstalling the Full Bottle Shut-Off Setting Up the Intake Line and StrainerSetting Up the Intake Line and Strainer Choosing BottlesPower Connections Power Connections12 V dc 12 V dc Connector Pin AssignmentsAuxiliary Connection Auxiliary Pin AssignmentsMA Input Contact ClosuresAuxiliary Connection Splitter Interface 20 mA Interface and Pulse Duration InputPage Keypad and Key Functions Tips and Techniques for Programming the Instrument Tips and Techniques for Programming the InstrumentBottle Setup Mode Section Setting the Time and DateSetting the Time and Date 1035 AM 24OCT00Section Explanations of Program Messages Data Logging Manual ModeManually Operating the Sample Pump Mode Measurements Triggered by Liquid Levels or FlowMeasurements Triggered by Liquid Levels or Flow Other Displayed Messages Section Programming the Sampler for Level ControlSection Flow Proportional Operation How to Calculate Pulses/CountsExample Flow Proportional OperationTherefore, the value for Intv = Cnts is Calculate t You want to collect 32 samples over an 8-hour period Page Removing and Opening the Controller Cleaning the SamplerRe-installing the Bottom Panel Re-installing the Bottom PanelCircuit Board Identification Section Internal Desiccant ModuleInternal Desiccant Module Memory Battery Internal Case Humidity IndicatorMotor/Gear Box Motor/Gear BoxOrdering Information for the U.S.A Repair ServiceContact Information for Europe Limitations Limited WarrantyLimitation of Remedies Page Description Part Number Parts and AccessoriesTray/Positioner Page Program Flow Chart Page Temperature MaintenanceLead-Acid Gel Cell Batteries ChargingNickel-Cadmium Battery Storage Recommendations Lead-Acid Battery Storage RecommendationsStorage Temperature Maximum Recommended Storage Time Appendix BItem Number Description Exploded View 1Exploded View Drawings Appendix C Exploded View 2 Appendix C Exploded View 3 Part Description Appendix C Exploded View 4Part Appendix C Exploded View 5 Item Number Description Part Number Appendix C Exploded View 6