4.11 Flow Proportional Operation

4.11.1 How to Calculate Pulses/Counts

4.11.1.1 Flow Proportional Sampling Intervals—Using External Pulses

The sampler is equipped to receive either a momentary dry contact closure or a +5 to +12 V dc pulse from a flow meter, where each pulse represents a known flow increment.

For example, if the flow meter is set to send a flow pulse for every 1000 gallons of measured flow and the sampler can count 1 to 9999 flow pulses between sample intervals, the sampler can be programmed to receive 10 pulses between samples to take one sample for every 10,000 gallons of liquid flow.

To determine the number of flow pulses between sample intervals, determine the total number of samples (represented by n in the following examples) to be collected, and the period of time over which they are to be collected. Also know the total flow, Q, during the sampling program. The following two-step method should be useful for determining the value for INTV = _ _ _ _ CNTS during setup.

1.Determine f, where f represents the flow increment between samples and n represents the total number of samples to be collected.

Q f = ----

n

2.Multiply the flow increment between samples, f, by the pulse frequency output of the flow meter (i.e. 1 pulse per 100 gallons, 1 pulse per 1,000 gallons, etc.). If the pulse frequency output of the flow meter is not known, consult the flow meter manufacturer.

Note: The final result is not necessarily a whole number. You must round off the result to the nearest whole number.

Example 1

You want to collect 35 samples over a 24-hour period. The total expected flow over this period is 235,000 gallons. The flow meter pulse frequency is

one pulse for every 100 gallons.

1.Determine the flow increment between samples:

Q 235,000 gallons

f = ---- = -----------------------------------------= 6,714 gallons/sample

n 35 samples

2. Multiply the flow increment, f, by the pulse frequency output of the flow meter.

6,714 gallons/sample ×

1 pulse

= 67.14 pulse/sample

100------------gallons------------------

Therefore, the value for INTV = _ _ _ _ CNTS is 67. (67.14 is rounded off to 67.)

Example 2

You want to collect 24 samples over an 8-hour period. The total expected flow over this 8-hour period is 85,000 gallons. The flow meter pulse frequency output is one pulse for every 50 gallons.

Page 28

 

Flow Proportional Operation

8544operations.fm

Image 28
Contents Sigma 900 Composite Sampler Page Table of Contents Table of Contents Controller Dimensions Pump/Controller HousingSample Volume Sample Volume RepeatabilityPage Interface Connectors Instrument DescriptionReceptacle Caps General InformationHumidity Indicator SectionFront Panel Keypad and DisplaySafety Information Safety InformationUse of Hazard Information Precautionary LabelsHazardous Locations Confined Space Entry PeligroDefinition of Confined Space Installation Selecting the Installation SiteUnpacking the Instrument Mounting the Sampler Attaching the Intake LineVinyl Intake Tubing Teflon-Lined Polyethylene Intake LineSetting Up the Intake Line and Strainer Installing the Full Bottle Shut-OffSetting Up the Intake Line and Strainer Choosing BottlesPower Connections Power Connections12 V dc Connector Pin Assignments 12 V dcAuxiliary Connection Auxiliary Pin AssignmentsMA Input Contact ClosuresAuxiliary Connection 20 mA Interface and Pulse Duration Input Splitter InterfacePage Keypad and Key Functions Tips and Techniques for Programming the Instrument Tips and Techniques for Programming the InstrumentSection Setting the Time and Date Bottle Setup ModeSetting the Time and Date 1035 AM 24OCT00Section Explanations of Program Messages Data Logging Manual ModeManually Operating the Sample Pump Mode Measurements Triggered by Liquid Levels or FlowMeasurements Triggered by Liquid Levels or Flow Section Programming the Sampler for Level Control Other Displayed MessagesHow to Calculate Pulses/Counts Section Flow Proportional OperationExample Flow Proportional OperationTherefore, the value for Intv = Cnts is Calculate t You want to collect 32 samples over an 8-hour period Page Cleaning the Sampler Removing and Opening the ControllerRe-installing the Bottom Panel Re-installing the Bottom PanelCircuit Board Identification Section Internal Desiccant ModuleInternal Desiccant Module Internal Case Humidity Indicator Memory BatteryMotor/Gear Box Motor/Gear BoxRepair Service Ordering Information for the U.S.AContact Information for Europe Limitations Limited WarrantyLimitation of Remedies Page Description Part Number Parts and AccessoriesTray/Positioner Page Program Flow Chart Page Maintenance TemperatureLead-Acid Gel Cell Batteries ChargingLead-Acid Battery Storage Recommendations Nickel-Cadmium Battery Storage RecommendationsStorage Temperature Maximum Recommended Storage Time Appendix BItem Number Description Exploded View 1Exploded View Drawings Appendix C Exploded View 2 Appendix C Exploded View 3 Part Description Appendix C Exploded View 4Part Appendix C Exploded View 5 Appendix C Exploded View 6 Item Number Description Part Number