2.0 CALIBRATION PROCEDURE

This section begins with a high-level overview of the calibration procedure for the M201E analyzer. Details are provided starting in Section 2.1:

First, take the time to read the M200E manual to familiarize yourself with the Chemiluminescence process. It is important to remember that the analyzer is merely measuring different levels of nitric oxide sample gas on three separate channels (TNx, NOx, NO). The ammonia and nitrogen dioxide concentrations are then calculated using this information.

Assemble the M201E analyzer according to the pneumatic configurations outlined in Figure 1-1. Remember: DO NOT pressurize the sample, span or zero ports (Zero/SPAN valves are optional) during calibration. Allow the analyzer to pull the gas through the system using the vacuum pump.

CAUTION!

If the presence of ozone is detected at any time, call Teledyne API Customer Service as soon as possible:

800-324-5190 or email: api-customerservice@teledyne.com

Next, zero the analyzer using an approved zero air source, such as, bottled zero air, nitrogen or zero air generated by a Teledyne-API M701 zero air source. This sets the zero offset for the three individual channels (TNx_OFFS, NOx_OFFS, NO_OFFS). Confirm that all displayed concentrations read zero.

Span the analyzer using bottled nitric oxide gas diluted to a level set to 80% of the expected sample range, preferably using a M702 Calibrator. This sets the slope for the three individual channels (TNx_SLOPE, NOx_SLOPE, NO_SLOPE). Confirm that displayed concentrations TNx , NOx and NO display the correct span concentration.

If you haven’t done so recently, conduct a GPT (gas phase titration) and confirm the molybdenum is functioning accordingly. Section 8.3.3 of the M200E manual outlines the GPT procedure. Section 7.0 of this manual consists of a service note for checking the molybdenum converter. There should be no reason to adjust the molybdenum efficiency factor on a new analyzer. If a diluted bottled of nitrogen dioxide gas is used to determine the molybdenum efficiency, allow enough time for the span value to stabilize. Nitrogen dioxide exhibits similar hold up issues as ammonia gas. Therefore, it may take a number of hours before the NOx reading stabilizes.

Span the analyzer using bottled ammonia gas diluted to a level set to 80% of the expected sample range, preferably using a M702 Calibrator. Note:

05206H DCN5910

15

Image 17
Contents Manual Addendum Model M201E Ammonia AnalyzerPage Name/Description About this ManualThis page intentionally left blank 05206H DCN5910 Safety Messages This page intentionally left blank 05206H DCN5910 Table of Contents Analog Output Manual Calibration List of TablesDo not pressurize the Sample/Span/Zero port above ambient M201E Ammonia AnalyzerTeledyne API Model M201E NH3 Analyzer Operator Manual M201E and M501 NH 3, Pneumatic Flow Theory of Operation Minimizing PMT Drift Purging the Reaction Cell Sample Filtration M201E Analyzer Specifications Calibration Procedure Step No Action Comment Zero Calibration Procedure Zero Gas Through the Sample PortZero Calibration Procedure Zero Gas Through Zero Port No Conv No Calibration Procedure no Gas Through the Sample PortNo Calibration Procedure no Gas through the Span Port Spanning the Analyzer with Ammonia GAS Press EXIT, Exit Confirming Ammonia Converter EfficiencyNH3 Calibration Procedure NH3 Gas Through The Sample Port NH3 Calibration Procedure NH3 Gas through the Span Port This page intentionally left blank 05206H DCN5910 Analog Output Data Default Settings Configurable Analog Output OverviewAnalog Output Pin-Outs Analog Output Current Loop Min/Max Analog Output Voltage Range Min/MaxAnalog Output Signal Type Selecting Auto or Manual Analog Output Calibration Auto Or Manual Analog Output Calibration Analog Output Group Calibration Setup For Calibrating Analog Output Signal Levels ZERO/SPAN Analog Output Signal Calibration TolerancesAnalog Output Manual Calibration Analog Output Recorder Offset Calibration Preventative Maintenance Schedule Maintenance ScheduleCatalytic Cartridge M501 NH3 Converter Rebuild Parts ListThermocouple Location Teledyne API Model M201E NH3 Analyzer Operator Manual Analyzer Flow Check This page intentionally left blank Alarms and Cautions M201E Functional Block Diagram Part Number Description M201E Spare Parts ListPCA, TEC DRIVER, PMT, M100/200, E Series ORING, 105M, 1MM W X 5 MM ID, Viton This page intentionally left blank 05206H DCN5910 None Calculating Moly Converter EfficienciesCalculating Moly Converter Efficiencies Moly Test Data Sheet

M201E specifications

The Teledyne M201E is a high-performance, compact oceanographic and environmental monitoring device designed for versatile underwater applications. Leveraging advanced technologies, the M201E presents a plethora of features appealing to researchers, engineers, and environmentalists seeking reliable data collection in challenging aquatic environments.

One of the standout characteristics of the M201E is its robust design, which allows it to operate in various underwater conditions, from shallow coastal regions to deep-sea environments. Built with durable materials, the device ensures long-term performance and resilience against corrosion, making it an ideal choice for long-term deployments.

The M201E is equipped with a range of sophisticated sensors, enabling it to collect comprehensive data on several environmental parameters. Key features include the ability to measure temperature, salinity, depth, and turbidity, among others. This multi-parameter capability allows for detailed assessments of underwater ecosystems and provides essential insights into vital oceanographic processes.

Another remarkable aspect of the M201E is its connectivity and data transmission technologies. It supports various communication protocols, enabling real-time data streaming to research teams, enhancing their ability to respond to changing conditions swiftly. The device can be integrated into larger networks of sensors, facilitating a more extensive monitoring system that collaborates and shares valuable data.

Moreover, the M201E is powered by advanced battery technologies that maximize operational longevity, allowing for extended missions without the need for frequent maintenance. Users can also benefit from intuitive user interfaces and software, streamlining the process of configuring the device, managing data, and analyzing results.

The M201E’s versatility is further demonstrated by its compatibility with various mounting options and its adaptability to different research or monitoring projects. Whether deployed by autonomous underwater vehicles (AUVs), buoys, or fixed platforms, the M201E meets diverse field requirements.

In conclusion, the Teledyne M201E stands out as a highly functional and reliable tool for underwater research. With its robust design, comprehensive sensing capabilities, advanced communications technologies, and ease of integration, it significantly contributes to our understanding of marine and freshwater environments. As the demand for precise and reliable environmental monitoring continues to grow, the M201E is poised to play a crucial role in the future of oceanographic research and environmental management.