Teledyne M201E manual Zero Calibration Procedure Zero Gas Through the Sample Port

Page 18

Calibration Procedure

Teledyne API Model M201E NH3 Analyzer Operator Manual

The first time bottled ammonia is connected to the gas dilution system, regulator/calibrator, the TNx reading may take a number of hours to stabilize.

2.1ZEROING THE ANALYZER

The analyzer can be zeroed by either applying zero air straight to the sample port or through the optional Zero/Span valves, if installed. It is important to remember that if the analyzer was previously sampling ammonia gas, prior to conducting a Zero calibration, it will take some period of time before the zero baseline is reached. The process of zeroing the analyzer consists of delivering dry zero air to the analyzer. The operator then manually zeros the TNx and NOx channels individually through the user interface. The following Tables outline the steps necessary to zero the analyzer.

Table 2-1. Zero Calibration Procedure – Zero Gas Through the SAMPLE Port

Important: Each channel (TNx and NOx) must be zeroed individually

Step No.

Action

Comment

 

 

 

1.

Press CAL

The M201E enters the calibrate mode from sample

 

 

mode. Confirm zero gas is flowing past the sample

 

 

port.

2.

Channel Selection

Press TNX or NOx then Press ENTR.

3.

Range Selection

Press LOW or HIGH range then Press ENTR.

 

 

Always calibrate the LOW range, section 3.0 below.

4.

NH3 STB Reading

Wait for the displayed stability reading to fall below

 

 

1.0 PPB. The lower the stability reading the better the

 

 

zero value.

5.

Press ZERO

If you change your mind after pressing ZERO, you

 

 

can still press EXIT without zeroing the instrument.

 

 

You don’t have to enter concentration values. The

 

 

analyzer knows to apply 0 PPB.

6.

Press ENTR

Pressing ENTR actually adjusts the Offset value for

 

 

the selected channel.

7.

Press EXIT

The M201E returns to sampling mode.

8.

Check

All displayed concentrations should read zero. The

 

Concentrations

offset values should be close to zero (-20 to 150 mv)

 

/Offsets

 

16

05206H DCN5910

Image 18
Contents Model M201E Ammonia Analyzer Manual AddendumPage About this Manual Name/DescriptionThis page intentionally left blank 05206H DCN5910 Safety Messages This page intentionally left blank 05206H DCN5910 Table of Contents List of Tables Analog Output Manual CalibrationM201E Ammonia Analyzer Do not pressurize the Sample/Span/Zero port above ambientTeledyne API Model M201E NH3 Analyzer Operator Manual M201E and M501 NH 3, Pneumatic Flow Theory of Operation Minimizing PMT Drift Purging the Reaction Cell Sample Filtration M201E Analyzer Specifications Calibration Procedure Zero Calibration Procedure Zero Gas Through the Sample Port Step No Action CommentZero Calibration Procedure Zero Gas Through Zero Port No Calibration Procedure no Gas Through the Sample Port No ConvNo Calibration Procedure no Gas through the Span Port Spanning the Analyzer with Ammonia GAS Confirming Ammonia Converter Efficiency Press EXIT, ExitNH3 Calibration Procedure NH3 Gas Through The Sample Port NH3 Calibration Procedure NH3 Gas through the Span Port This page intentionally left blank 05206H DCN5910 Configurable Analog Output Overview Analog Output Data Default SettingsAnalog Output Pin-Outs Analog Output Voltage Range Min/Max Analog Output Current Loop Min/MaxAnalog Output Signal Type Selecting Auto or Manual Analog Output Calibration Auto Or Manual Analog Output Calibration Analog Output Group Calibration ZERO/SPAN Analog Output Signal Calibration Tolerances Setup For Calibrating Analog Output Signal LevelsAnalog Output Manual Calibration Analog Output Recorder Offset Calibration Maintenance Schedule Preventative Maintenance ScheduleM501 NH3 Converter Rebuild Parts List Catalytic CartridgeThermocouple Location Teledyne API Model M201E NH3 Analyzer Operator Manual Analyzer Flow Check This page intentionally left blank Alarms and Cautions M201E Functional Block Diagram M201E Spare Parts List Part Number DescriptionPCA, TEC DRIVER, PMT, M100/200, E Series ORING, 105M, 1MM W X 5 MM ID, Viton This page intentionally left blank 05206H DCN5910 Calculating Moly Converter Efficiencies NoneCalculating Moly Converter Efficiencies Moly Test Data Sheet

M201E specifications

The Teledyne M201E is a high-performance, compact oceanographic and environmental monitoring device designed for versatile underwater applications. Leveraging advanced technologies, the M201E presents a plethora of features appealing to researchers, engineers, and environmentalists seeking reliable data collection in challenging aquatic environments.

One of the standout characteristics of the M201E is its robust design, which allows it to operate in various underwater conditions, from shallow coastal regions to deep-sea environments. Built with durable materials, the device ensures long-term performance and resilience against corrosion, making it an ideal choice for long-term deployments.

The M201E is equipped with a range of sophisticated sensors, enabling it to collect comprehensive data on several environmental parameters. Key features include the ability to measure temperature, salinity, depth, and turbidity, among others. This multi-parameter capability allows for detailed assessments of underwater ecosystems and provides essential insights into vital oceanographic processes.

Another remarkable aspect of the M201E is its connectivity and data transmission technologies. It supports various communication protocols, enabling real-time data streaming to research teams, enhancing their ability to respond to changing conditions swiftly. The device can be integrated into larger networks of sensors, facilitating a more extensive monitoring system that collaborates and shares valuable data.

Moreover, the M201E is powered by advanced battery technologies that maximize operational longevity, allowing for extended missions without the need for frequent maintenance. Users can also benefit from intuitive user interfaces and software, streamlining the process of configuring the device, managing data, and analyzing results.

The M201E’s versatility is further demonstrated by its compatibility with various mounting options and its adaptability to different research or monitoring projects. Whether deployed by autonomous underwater vehicles (AUVs), buoys, or fixed platforms, the M201E meets diverse field requirements.

In conclusion, the Teledyne M201E stands out as a highly functional and reliable tool for underwater research. With its robust design, comprehensive sensing capabilities, advanced communications technologies, and ease of integration, it significantly contributes to our understanding of marine and freshwater environments. As the demand for precise and reliable environmental monitoring continues to grow, the M201E is poised to play a crucial role in the future of oceanographic research and environmental management.