Teledyne API Model M201E NH3 Analyzer Operator Manual

signal. This process improves zero baseline stability by minimizing the effect of PMT sensor drift.

1.1.2Purging the Reaction Cell

As with many chemical reactions the conversion of ammonia in the presence of other oxides of nitrogen is complicated. It is important to note that the valve DWELL time for an AZERO measurement has a default setting of 8 seconds. Shortening this value may not allow enough time to properly purge the reaction cell of excess nitric oxide from the previous measurements.

In the molybdenum converter operating at 315oC the following significant reactions are taking place:

Mo + NO2 MoO3 + NO ~100% Efficiency

The M501 NH3, ammonia converter operates at 825oC. At this high temperature, several reactions occur:

NO NO

Loss = ~ 3%

NO2 NO

Efficiency = ~97%

NH3

NO

Efficiency = ~97%

NH3

NO2

Efficiency = ~5%

As can be seen from the above reactions, the calculation of the ammonia concentration and overall calibration of the instrument must be done carefully, if accurate ammonia concentrations are to be measured.

1.2SPECIAL CONSIDERATIONS FOR AMMONIA MEASUREMENT

Ammonia is a difficult gas to measure due to its chemical characteristics. The gas tends to adsorb onto surfaces and diffuse into many materials. The following precautions should be observed when designing ammonia sampling systems and connecting them to the M201E analyzer:

Do NOT use copper tubing or fittings designed for household plumbing.

Use ONLY Chromatography grade (cleaned, passivated) stainless steel tubing.

Use ONLY Glass tubing for sample inlet manifold.

ALWAYS keep the tubing from the M702SS calibrator to the analyzer as short as possible. USE stainless steel tubing throughout, especially from the ammonia calibration bottle to the calibrator.

If possible HEAT the sample line and DRY the sample gas.

Sample filters Will cause response delays on both the rise and fall of the sample signal.

12

05206H DCN5910

Image 14
Contents Model M201E Ammonia Analyzer Manual AddendumPage About this Manual Name/DescriptionThis page intentionally left blank 05206H DCN5910 Safety Messages This page intentionally left blank 05206H DCN5910 Table of Contents List of Tables Analog Output Manual CalibrationM201E Ammonia Analyzer Do not pressurize the Sample/Span/Zero port above ambientTeledyne API Model M201E NH3 Analyzer Operator Manual M201E and M501 NH 3, Pneumatic Flow Theory of Operation Minimizing PMT Drift Purging the Reaction Cell Sample Filtration M201E Analyzer Specifications Calibration Procedure Zero Calibration Procedure Zero Gas Through the Sample Port Step No Action CommentZero Calibration Procedure Zero Gas Through Zero Port No Calibration Procedure no Gas Through the Sample Port No ConvNo Calibration Procedure no Gas through the Span Port Spanning the Analyzer with Ammonia GAS Confirming Ammonia Converter Efficiency Press EXIT, ExitNH3 Calibration Procedure NH3 Gas Through The Sample Port NH3 Calibration Procedure NH3 Gas through the Span Port This page intentionally left blank 05206H DCN5910 Configurable Analog Output Overview Analog Output Data Default SettingsAnalog Output Pin-Outs Analog Output Voltage Range Min/Max Analog Output Current Loop Min/MaxAnalog Output Signal Type Selecting Auto or Manual Analog Output Calibration Auto Or Manual Analog Output Calibration Analog Output Group Calibration ZERO/SPAN Analog Output Signal Calibration Tolerances Setup For Calibrating Analog Output Signal LevelsAnalog Output Manual Calibration Analog Output Recorder Offset Calibration Maintenance Schedule Preventative Maintenance ScheduleM501 NH3 Converter Rebuild Parts List Catalytic CartridgeThermocouple Location Teledyne API Model M201E NH3 Analyzer Operator Manual Analyzer Flow Check This page intentionally left blank Alarms and Cautions M201E Functional Block Diagram M201E Spare Parts List Part Number DescriptionPCA, TEC DRIVER, PMT, M100/200, E Series ORING, 105M, 1MM W X 5 MM ID, Viton This page intentionally left blank 05206H DCN5910 Calculating Moly Converter Efficiencies NoneCalculating Moly Converter Efficiencies Moly Test Data Sheet

M201E specifications

The Teledyne M201E is a high-performance, compact oceanographic and environmental monitoring device designed for versatile underwater applications. Leveraging advanced technologies, the M201E presents a plethora of features appealing to researchers, engineers, and environmentalists seeking reliable data collection in challenging aquatic environments.

One of the standout characteristics of the M201E is its robust design, which allows it to operate in various underwater conditions, from shallow coastal regions to deep-sea environments. Built with durable materials, the device ensures long-term performance and resilience against corrosion, making it an ideal choice for long-term deployments.

The M201E is equipped with a range of sophisticated sensors, enabling it to collect comprehensive data on several environmental parameters. Key features include the ability to measure temperature, salinity, depth, and turbidity, among others. This multi-parameter capability allows for detailed assessments of underwater ecosystems and provides essential insights into vital oceanographic processes.

Another remarkable aspect of the M201E is its connectivity and data transmission technologies. It supports various communication protocols, enabling real-time data streaming to research teams, enhancing their ability to respond to changing conditions swiftly. The device can be integrated into larger networks of sensors, facilitating a more extensive monitoring system that collaborates and shares valuable data.

Moreover, the M201E is powered by advanced battery technologies that maximize operational longevity, allowing for extended missions without the need for frequent maintenance. Users can also benefit from intuitive user interfaces and software, streamlining the process of configuring the device, managing data, and analyzing results.

The M201E’s versatility is further demonstrated by its compatibility with various mounting options and its adaptability to different research or monitoring projects. Whether deployed by autonomous underwater vehicles (AUVs), buoys, or fixed platforms, the M201E meets diverse field requirements.

In conclusion, the Teledyne M201E stands out as a highly functional and reliable tool for underwater research. With its robust design, comprehensive sensing capabilities, advanced communications technologies, and ease of integration, it significantly contributes to our understanding of marine and freshwater environments. As the demand for precise and reliable environmental monitoring continues to grow, the M201E is poised to play a crucial role in the future of oceanographic research and environmental management.