Teledyne M201E manual NH3 Calibration Procedure NH3 Gas Through The Sample Port

Page 24

Calibration Procedure

Teledyne API Model M201E NH3 Analyzer Operator Manual

Table 2-6. NH3 Calibration Procedure – NH3 Gas Through The SAMPLE Port

Step No.

Action

Comment

 

 

 

 

1.

Generate a known

Monitor the TNx concentration until it has stabilized;

 

level of NH3 gas

this signifies that both the NOx and NH3 gases have

 

using a calibration

stabilized. This could range from 20 minutes to

 

gas source

several hours depending on whether the analyzer /

 

 

calibration system has recently been subjected to

 

 

ammonia gas.

 

2.

NH3 Conversion

When the TNx channel is stable, note the

 

efficiency

concentration of NH3 displayed. The calculation for

 

calculation

the NH3 Conversion efficiency is:

 

 

 

(Displayed NH3 concentration) divided by

 

 

(Delivered NH3 concentration)

 

 

 

Example: The operator is delivering 450 PPB of NH3.

 

 

The displayed concentration for NH3 is 440 PPB then

 

 

the efficiency factor = 440/450 = 0.9777

3.

Press CAL

The M201E enters the calibrate mode from Sample

 

 

mode.

 

4.

Channel Selection

Press TNX then Press ENTR.

 

5.

Range Selection

Press LOW range then Press ENTR.

6.

Press CONC

Since TNx was chosen previously, then the following

 

 

will be displayed:

 

 

 

TNx CONV

Exit

 

 

Press CONV enter 0.9777, Press ENTR

 

 

Press EXIT, EXIT

 

Note: In the example outlined above, the displayed ammonia concentration is being increased by 2.27%. Newer M501 NH3, converters should have efficiency values very close to 1.00. Depending on the operator’s acceptable level of error, it may not be necessary to change the efficiency factor from 1.00. Therefore, calibration with nitric oxide gas is all that is required.

The efficiency factor may also be changed by entering the VARS MENU and editing NH3_CE_FACTOR1 = 1.0000.

22

05206H DCN5910

Image 24
Contents Model M201E Ammonia Analyzer Manual AddendumPage About this Manual Name/DescriptionThis page intentionally left blank 05206H DCN5910 Safety Messages This page intentionally left blank 05206H DCN5910 Table of Contents List of Tables Analog Output Manual CalibrationM201E Ammonia Analyzer Do not pressurize the Sample/Span/Zero port above ambientTeledyne API Model M201E NH3 Analyzer Operator Manual M201E and M501 NH 3, Pneumatic Flow Theory of Operation Minimizing PMT Drift Purging the Reaction Cell Sample Filtration M201E Analyzer Specifications Calibration Procedure Zero Calibration Procedure Zero Gas Through the Sample Port Step No Action CommentZero Calibration Procedure Zero Gas Through Zero Port No Calibration Procedure no Gas Through the Sample Port No ConvNo Calibration Procedure no Gas through the Span Port Spanning the Analyzer with Ammonia GAS Confirming Ammonia Converter Efficiency Press EXIT, ExitNH3 Calibration Procedure NH3 Gas Through The Sample Port NH3 Calibration Procedure NH3 Gas through the Span Port This page intentionally left blank 05206H DCN5910 Configurable Analog Output Overview Analog Output Data Default SettingsAnalog Output Pin-Outs Analog Output Voltage Range Min/Max Analog Output Current Loop Min/MaxAnalog Output Signal Type Selecting Auto or Manual Analog Output Calibration Auto Or Manual Analog Output Calibration Analog Output Group Calibration ZERO/SPAN Analog Output Signal Calibration Tolerances Setup For Calibrating Analog Output Signal LevelsAnalog Output Manual Calibration Analog Output Recorder Offset Calibration Maintenance Schedule Preventative Maintenance ScheduleM501 NH3 Converter Rebuild Parts List Catalytic CartridgeThermocouple Location Teledyne API Model M201E NH3 Analyzer Operator Manual Analyzer Flow Check This page intentionally left blank Alarms and Cautions M201E Functional Block Diagram M201E Spare Parts List Part Number DescriptionPCA, TEC DRIVER, PMT, M100/200, E Series ORING, 105M, 1MM W X 5 MM ID, Viton This page intentionally left blank 05206H DCN5910 Calculating Moly Converter Efficiencies NoneCalculating Moly Converter Efficiencies Moly Test Data Sheet

M201E specifications

The Teledyne M201E is a high-performance, compact oceanographic and environmental monitoring device designed for versatile underwater applications. Leveraging advanced technologies, the M201E presents a plethora of features appealing to researchers, engineers, and environmentalists seeking reliable data collection in challenging aquatic environments.

One of the standout characteristics of the M201E is its robust design, which allows it to operate in various underwater conditions, from shallow coastal regions to deep-sea environments. Built with durable materials, the device ensures long-term performance and resilience against corrosion, making it an ideal choice for long-term deployments.

The M201E is equipped with a range of sophisticated sensors, enabling it to collect comprehensive data on several environmental parameters. Key features include the ability to measure temperature, salinity, depth, and turbidity, among others. This multi-parameter capability allows for detailed assessments of underwater ecosystems and provides essential insights into vital oceanographic processes.

Another remarkable aspect of the M201E is its connectivity and data transmission technologies. It supports various communication protocols, enabling real-time data streaming to research teams, enhancing their ability to respond to changing conditions swiftly. The device can be integrated into larger networks of sensors, facilitating a more extensive monitoring system that collaborates and shares valuable data.

Moreover, the M201E is powered by advanced battery technologies that maximize operational longevity, allowing for extended missions without the need for frequent maintenance. Users can also benefit from intuitive user interfaces and software, streamlining the process of configuring the device, managing data, and analyzing results.

The M201E’s versatility is further demonstrated by its compatibility with various mounting options and its adaptability to different research or monitoring projects. Whether deployed by autonomous underwater vehicles (AUVs), buoys, or fixed platforms, the M201E meets diverse field requirements.

In conclusion, the Teledyne M201E stands out as a highly functional and reliable tool for underwater research. With its robust design, comprehensive sensing capabilities, advanced communications technologies, and ease of integration, it significantly contributes to our understanding of marine and freshwater environments. As the demand for precise and reliable environmental monitoring continues to grow, the M201E is poised to play a crucial role in the future of oceanographic research and environmental management.