Chapter 10 System Checkout Procedures
139
This chapter details the test procedures that should be performed by
the clinician prior to connecting the device to the patient. Test the
Active PAP, Active Flow, and Passive circuit types if you want to do a
complete checkout on the device. The tests should be performed as
described in order to verify proper operation of the device. Some of
the procedures in this chapter require you to change settings on the
device. If you are not familiar with the procedures for accomplishing
this, please refer to Chapter 4, Ventilator Setup and Chapter 5, Viewing
and Changing Settings.

Tools Required

Universal Porting Block

Active PAP Exhalation Device (with Tube Adapter

connected)

Active Flow Exhalation Device with Flow Sensor

Passive Exhalation Device

Test Lung

Small Flat Head Screwdriver

Visual Inspection

1. Verify that the enclosure is not broken and that all applicable
screws are in place.
Trilogy200
clinical manual
10. System Checkout Procedures
Note: The actual circuit
conguration to be used on
the patient should be used to
perform the system checkout
procedure.
WARNING
If you notice any unexplained
changes in the performance
of the device, if it is making
unusual sounds, if the device
or detachable battery are
dropped, if water is spilled into
the enclosure, or if the enclosure
is cracked or broken, discontinue
use and contact Philips
Respironics or an authorized
service center for service.