3

Warnings and Cautions

Caution: U.S. federal law restricts this device to sale by or on the order of a physician.

Warnings

A warning indicates the possibility of injury to the user or operator.

Patient Monitoring

Prior to placing a patient on the ventilator, a clinical assessment should

 

be performed to determine:

 

The device alarm settings

 

Needed alternative ventilation equipment

 

If an alternative monitor (i.e., an alarming Pulse Oximeter or

 

Respiratory Monitor) should be used

Alternative

For ventilator dependent patients, always have alternate ventilation

Ventilation

equipment, such as a back-up ventilator, manual resuscitator, or similar

 

device, available.

 

Ventilator dependent patients should be continuously monitored by

 

qualified personnel. These personnel should be prepared to provide

 

alternate therapy in the event of ventilator failure or inoperative

 

equipment.

Patient Disconnect

For ventilator dependent patients, do not rely on any single alarm

Protection

to detect a circuit disconnect condition. The Low Tidal Volume, Low

 

Minute Ventilation, Low Respiratory Rate, and Apnea alarms should

 

be used in conjunction with the Circuit Disconnect and Low Peak

 

Inspiratory Pressure alarms.

 

Test the operation of the circuit disconnect function daily and

 

whenever a change is made to the patient circuit. An increase in circuit

 

resistance can prevent proper operation of some alarms.

 

Speaking valves, Heat Moisture Exchangers (HMEs), and filters create

 

additional circuit resistance and may affect the performance of alarms

 

chosen for circuit disconnect protection.

 

Do not set the Low Peak Inspiratory Pressure alarm too low, or the

 

system may not detect large circuit leaks or a patient disconnect.

Chapter 1 Introduction

Page 14
Image 14
Philips Trilogy 200 manual Patient Monitoring, Alternative, Ventilation, Patient Disconnect, Protection