6

Remote Alarms

When using a remote alarm, make sure you fully test the remote alarm

 

connector and cable by verifying that:

 

Annunciated alarms on the ventilator are also

 

annunciated on the remote alarm.

 

Disconnecting the remote alarm cable from the

 

ventilator or from the remote alarm results in an alarm

 

notification at the remote alarm.

 

The remote alarm should be tested daily.

Oxygen

When administering fixed-flow supplemental oxygen, the oxygen

 

concentration may not be constant. The inspired oxygen concentration

 

will vary, depending on the pressures, patient flows and circuit leak.

 

Substantial leaks may reduce the inspired oxygen concentration to less

 

than the expected value. Appropriate patient monitoring should be

 

used, as medically indicated, such as an alarming pulse oximeter.

 

This device DOES NOT alarm for loss of the low flow oxygen supply.

 

This device is intended to be connected to a low flow (0-15 l/min)

 

oxygen source such as an oxygen concentrator or other oxygen source

 

equipped with a pressure regulator (set to 50 PSI or less) and a flow

 

regulator/meter.

 

Do not connect the device to an unregulated or high pressure oxygen

 

source.

 

The device may result in incorrect flow and tidal volume measurements

 

and improper operation of related alarms if you add low flow oxygen

 

directly into the patient circuit or mask instead of using the oxygen

 

inlet on the back of the ventilator.

 

Oxygen supports combustion. Oxygen should not be used while

 

smoking or in the presence of an open flame.

 

If oxygen is used with the device, the oxygen flow must be turned

 

off when the device is not in use.Explanation of the Warning:

 

When the device is not in operation and the oxygen flow is left on,

 

oxygen delivered into the tubing may accumulate within the device’s

 

enclosure.

Fire or Explosion

The ventilator should not be operated in the presence of flammable

 

gasses. This could cause a fire or explosion.

Trilogy200 clinical manual

Page 17
Image 17
Philips Trilogy 200 manual Remote Alarms, Oxygen, Fire or Explosion