59

 

 

Proximal

Universal Porting Block

Exhalation Valve Port

Pressure

Trilogy Universal Active

Line

 

 

PAP Tube Adapter

 

 

 

Proximal

 

 

 

Pressure

 

 

 

Port

Exhalation Valve Line

Exhalation Valve Port

 

 

 

 

on Universal Porting

 

 

 

Block

Connecting an Active Device with Proximal Pressure

Refer to the instructions included with the Active Exhalation Device for more information.

c.If your circuit type is Active Flow:

1.Connect the Flow Sensor to an active exhalation device as shown below.

Flow Sensor

2.Connect the other end of the flexible tubing to the active exhalation device.

A.Connect the active exhalation device to the flexible tubing that attaches to the ventilator.

B.Connect the flow sensor to the patient interface (e.g., the tracheostomy tube).

Connecting the Flow Sensor

Note: If the active exhalation device has the proximal pressure line connected when you are setting up an active flow circuit type, remove the proximal pressure line and cap the proximal pressure port before attaching to the flexible tubing.

Note: Make sure that components marked with an arrow are oriented properly.

Chapter 4 Ventilator Setup

Page 70