Connecting an Active Device with Proximal Pressure
Refer to the instructions included with the Active Exhalation Device for more information.
c.If your circuit type is Active Flow:
1.Connect the Flow Sensor to an active exhalation device as shown below.
Flow Sensor
2.Connect the other end of the flexible tubing to the active exhalation device.
A.Connect the active exhalation device to the flexible tubing that attaches to the ventilator.
B.Connect the flow sensor to the patient interface (e.g., the tracheostomy tube).
Connecting the Flow Sensor
Note: If the active exhalation device has the proximal pressure line connected when you are setting up an active flow circuit type, remove the proximal pressure line and cap the proximal pressure port before attaching to the flexible tubing.
Note: Make sure that components marked with an arrow are oriented properly.