132

Issue

Why It Happened

What To Do

Dryness

The air is too dry. This only

Increase the room humidity.

The patient has throat or

applies to non-invasive

Consult with the patient’s health

nose dryness.

therapy and will not occur

care professional about using

 

during invasive therapy.

a humidifier with the device. If

 

 

using a humidifier, refer to the

 

 

instructions included with the

 

 

humidifier to make sure it is

 

 

working properly.

Sinus or Ear Pain

The patient may have a sinus

Tell the patient to stop using the

The patient experiences

or middle ear infection. This

device and contact their health

nasal, sinus, or ear pain.

only applies to non-invasive

care professional.

 

therapy and will not occur

 

 

during invasive therapy.

 

Patient Tubing Condensation

Moisture added to the circuit

Use an appropriate method of

There is condensation in the

from a heated humidifier may

water management, such as a

circuit tubing or exhalation

condense. This condensation is

water trap when using a heated

valve lines.

typically referred to as“rain-out.”

humidifier.

Exhalation Valve or Flow

Excessive humidity (from

Remove tube(s) with

Tube Condensation (Active

the heated humidifier, the

condensation: the exhalation

Flow Circuit only)

patient, or the cleaning

valve line must be removed from

There is condensation

process) condensed in the

both ends of the circuit; the flow

in the proximal flow or

tubing or exhalation valve

tube lines must be removed

exhalation valve lines, and

diaphragm. Droplets of

from the ventilator side. The

the condensation is causing

water affected the ability of

tubes are permanently attached

the system to alarm or not

the ventilator to measure

to the flow element, so the flow

properly trigger on patient

proximal patient flow and

element must be removed from

effort.

pressure and trigger on a

the patient circuit. If there is

 

patient breath.

water built up in the exhalation

 

 

diaphragm, disassemble the

 

 

exhalation valve and clean/dry

 

 

the assembly. Clear the flow

 

 

tubes of water droplets using

 

 

a low flow air source and re-

 

 

assemble the circuit.

Trilogy200 clinical manual

Page 143
Image 143
Philips Trilogy 200 manual 132, Dryness, Sinus or Ear Pain, Exhalation Valve or Flow, Tube Condensation Active