Alarm | Priority | Audible | Visual (Alarm | Device | User/Clinician Action |
|
|
| Indicator Button | Action |
|
|
|
| and Display) |
|
|
High Internal | High | • • • | • • | Red flashing button; | Continues | User: Press Alarm Indicator/ |
|
| |||||
Oxygen |
|
|
| “High Internal | to operate | Audio Pause button to |
|
|
|
| Oxygen” message | when internal | silence alarm. Or, press Left |
|
|
|
|
| oxygen | (Reset) button to reset alarm. |
|
|
|
|
| concentration | Disconnect supplemental |
|
|
|
|
| reaches | oxygen supply from the |
|
|
|
|
| 5% above | ventilator and contact your |
|
|
|
|
| ambient | health care professional. |
|
|
|
|
| levels. | Clinician: Disconnect |
|
|
|
|
|
| |
|
|
|
|
|
| supplemental oxygen from |
|
|
|
|
|
| device. Check external |
|
|
|
|
|
| oxygen connection. If |
|
|
|
|
|
| problem continues, have |
|
|
|
|
|
| device serviced. |
Circuit | High | • • • | • • | Red flashing | button; | Operates | User: Press Alarm Indicator/ |
|
| ||||||
Disconnect |
|
|
| “Circuit Disconnect” |
| Audio Pause button to | |
|
|
|
| message |
|
| silence alarm. Or, press Left |
|
|
|
|
|
|
| (Reset) button to reset alarm. |
|
|
|
|
|
|
| Check breathing circuit and |
|
|
|
|
|
|
| reconnect it if it has become |
|
|
|
|
|
|
| disconnected, or fix the leak. |
|
|
|
|
|
|
| If alarm continues to occur, |
|
|
|
|
|
|
| contact your health care |
|
|
|
|
|
|
| professional. If device will not |
|
|
|
|
|
|
| exit circuit disconnect, switch |
|
|
|
|
|
|
| to an alternate source of |
|
|
|
|
|
|
| ventilation. |
|
|
|
|
|
|
| Clinician: Reconnect tubing |
|
|
|
|
|
|
| or fix leak. If alarm continues, |
|
|
|
|
|
|
| have device serviced. |
Apnea | High | • • • | • • | Red flashing | button; | Operates | User: Press Alarm Indicator/ |
|
| ||||||
|
|
|
| “Apnea” message |
| Audio Pause button to | |
|
|
|
|
|
|
| silence alarm. Or, press Left |
|
|
|
|
|
|
| (Reset) button to reset alarm. |
|
|
|
|
|
|
| Continue using device. If |
|
|
|
|
|
|
| using an Active PAP or Active |
|
|
|
|
|
|
| Flow circuit, check proximal |
|
|
|
|
|
|
| pressure line to ensure it is |
|
|
|
|
|
|
| not pinched or has water |
|
|
|
|
|
|
| condensation (refer to the |
|
|
|
|
|
|
| Troubleshooting section |
|
|
|
|
|
|
| – tubing condensation). If |
|
|
|
|
|
|
| alarm continues, contact your |
|
|
|
|
|
|
| health care professional. |
|
|
|
|
|
|
| Clinician: Verify patient |
|
|
|
|
|
|
| status. |