110

Alarm

Priority

Audible

Visual (Alarm

Device

User/Clinician Action

 

 

 

Indicator Button

Action

 

 

 

 

and Display)

 

 

High Internal

High

• • •

• •

Red flashing button;

Continues

User: Press Alarm Indicator/

 

 

Oxygen

 

 

 

“High Internal

to operate

Audio Pause button to

 

 

 

 

Oxygen” message

when internal

silence alarm. Or, press Left

 

 

 

 

 

oxygen

(Reset) button to reset alarm.

 

 

 

 

 

concentration

Disconnect supplemental

 

 

 

 

 

reaches

oxygen supply from the

 

 

 

 

 

5% above

ventilator and contact your

 

 

 

 

 

ambient

health care professional.

 

 

 

 

 

levels.

Clinician: Disconnect

 

 

 

 

 

 

 

 

 

 

 

 

supplemental oxygen from

 

 

 

 

 

 

device. Check external

 

 

 

 

 

 

oxygen connection. If

 

 

 

 

 

 

problem continues, have

 

 

 

 

 

 

device serviced.

Circuit

High

• • •

• •

Red flashing

button;

Operates

User: Press Alarm Indicator/

 

 

Disconnect

 

 

 

“Circuit Disconnect”

 

Audio Pause button to

 

 

 

 

message

 

 

silence alarm. Or, press Left

 

 

 

 

 

 

 

(Reset) button to reset alarm.

 

 

 

 

 

 

 

Check breathing circuit and

 

 

 

 

 

 

 

reconnect it if it has become

 

 

 

 

 

 

 

disconnected, or fix the leak.

 

 

 

 

 

 

 

If alarm continues to occur,

 

 

 

 

 

 

 

contact your health care

 

 

 

 

 

 

 

professional. If device will not

 

 

 

 

 

 

 

exit circuit disconnect, switch

 

 

 

 

 

 

 

to an alternate source of

 

 

 

 

 

 

 

ventilation.

 

 

 

 

 

 

 

Clinician: Reconnect tubing

 

 

 

 

 

 

 

or fix leak. If alarm continues,

 

 

 

 

 

 

 

have device serviced.

Apnea

High

• • •

• •

Red flashing

button;

Operates

User: Press Alarm Indicator/

 

 

 

 

 

 

“Apnea” message

 

Audio Pause button to

 

 

 

 

 

 

 

silence alarm. Or, press Left

 

 

 

 

 

 

 

(Reset) button to reset alarm.

 

 

 

 

 

 

 

Continue using device. If

 

 

 

 

 

 

 

using an Active PAP or Active

 

 

 

 

 

 

 

Flow circuit, check proximal

 

 

 

 

 

 

 

pressure line to ensure it is

 

 

 

 

 

 

 

not pinched or has water

 

 

 

 

 

 

 

condensation (refer to the

 

 

 

 

 

 

 

Troubleshooting section

 

 

 

 

 

 

 

– tubing condensation). If

 

 

 

 

 

 

 

alarm continues, contact your

 

 

 

 

 

 

 

health care professional.

 

 

 

 

 

 

 

Clinician: Verify patient

 

 

 

 

 

 

 

status.

Trilogy200 clinical manual

Page 121
Image 121
Philips Trilogy 200 manual 110