Trilogy200
clinical manual
129
8. Troubleshooting
The ventilator is designed to respond to most issues with an appropriate alarm or information message. These messages appear on the display screen. Chapter 6 describes many of the alarms and what action you should take when the alarms occur.
This chapter provides additional troubleshooting information for other common issues you may have.
Issue | Why It Happened | What To Do |
Device does not turn on. | The AC power cord is not | Plug the AC power cord into the |
Nothing happens when you | plugged in and the internal | device and a wall outlet that is |
press Start/Stop to begin | battery is not charged. | not controlled by a switch. This |
therapy. The audible indicator | Note: The device requires AC | powers the system and charges |
does not sound and the | power to charge the internal | the internal battery. If the device |
backlights on the buttons do | battery when a new device is | still does not turn on, contact an |
not light. | powered up for the first time. | authorized service representative |
|
| or Philips Respironics to have the |
|
| device serviced. Please have the |
|
| model number and serial number |
|
| ready when you call. |
Chapter 8 Troubleshooting