4.Connect the patient circuit. Fit the ventilator-end of the flexible tubing onto the bacteria filter.

Installing a New Bacteria

Filter

Preventive Maintenance

Your new Philips Respironics Trilogy ventilator includes a blank PM Service Due label affixed to the bottom of your device. The label is blank, which allows you to document when service is due. Service is due every 10,000 hours or 24 months, whichever comes first depending on the usage of the device.

Service Date

Service Hours

Preventive Maintenance Sticker

You will need to record both the date and hours on the blank label to maximize your in-service interval. Record the service date as 24 months from the date first placed on a patient. Record the hours as 10,000 hours for initial usage.

Use the device blower hours to determine when service is required. The device blower hours are listed in the device software in the information menu.

Trilogy200 clinical manual

Page 137
Image 137