Trilogy200 clinical manual
126
4. Connect the patient circuit. Fit the ventilator-end of the exible
tubing onto the bacteria lter.
Preventive Maintenance
Your new Philips Respironics Trilogy ventilator includes a blank PM
Service Due label axed to the bottom of your device. The label is
blank, which allows you to document when service is due. Service
is due every 10,000 hours or 24 months, whichever comes rst
depending on the usage of the device.

Service Date

Service Hours

You will need to record both the date and hours on the blank label
to maximize your in-service interval. Record the service date as 24
months from the date rst placed on a patient. Record the hours as
10,000 hours for initial usage.
Use the device blower hours to determine when service is required.
The device blower hours are listed in the device software in the
information menu.
Installing a New Bacteria
Filter
Preventive Maintenance
Sticker