Chapter 4 Ventilator Setup
49
This chapter provides instructions on how to assemble the ventilator.
It includes the following setup information:

Properly positioning the device

Installing the air lter

Supplying power to the device

Connecting the breathing circuit

Connecting a water trap (optional)

Connecting a remote alarm (optional)

WARNING
Ventilator dependent patients
should be continuously
monitored by qualied
personnel. These personnel
should be prepared to provide
alternate therapy in the event of
ventilator failure or inoperative
equipment.
WARNING
Do not use the ventilator on a
patient until a system checkout
has been performed. See
Chapter 10 of this manual.
WARNING
For ventilator dependent
patients, always have alternate
ventilation equipment, such as
a back-up ventilator, manual
resuscitator, or similar device,
available.
CAUTION
Do not operate the ventilator at
temperatures below 5° C (41° F)
or above 40° C (104° F).
Trilogy200
clinical manual
4. Ventilator Setup