Patient Circuit | Exhalation valves, patient circuits, and water traps are shipped clean, |
| not sterile. Cleaning and disinfection of these parts should follow |
| individual institution processes and conform to guidelines provided by |
| Philips Respironics with each accessory. |
External DC Power | Do not use the same external battery to operate both the ventilator |
| and any other equipment such as power chairs. |
| An external battery should only be connected to the ventilator using |
| the Philips Respironics Trilogy External Battery Cable. This cable is |
| fused, |
| to a standard |
| cable may cause improper operation of the ventilator. |
| The ventilator should only be connected to an automotive electrical |
| system using the Philips Respironics Trilogy Automotive Adapter |
| (when available). This adapter is fused, filtered, and designed for safe |
| connection to a standard automotive electrical system. Use of any |
| other adapter or cable may cause improper operation of the ventilator. |
| Do not operate the ventilator from a car electrical system when |
| starting the vehicle or |
| during starting may cause improper operation of the ventilator. |
Electrostatic | Do not use antistatic or conductive hoses or conductive patient tubing |
Discharge (ESD) | with the device. |
Notes
•• This product does not contain natural latex rubber or dry natural rubber in patient or operator accessible areas or in the air path or breathing circuit.
Trilogy200 clinical manual