Pressure Control (PC) Mode

In Pressure Control (PC) mode, the device delivers bi-level pressure support. This mode delivers Assist and Mandatory breaths. This mode is identical to S/T mode, except that all breaths have a fixed inspiratory time.

IPAP = 26 cm H2O

EPAP = 6 cm H2O

Rate = 10 BPM

PS = IPAP-EPAP = 20 cm H2O

PRESSURE

2= Mandatory Breath

3= Assist Breath

6 seconds

 

6 seconds

 

6 seconds

26

20

10

 

 

 

 

PC Mode

3

3

2

3

 

6

 

 

 

 

 

 

 

Enter PC mode

 

 

 

 

 

5

10

15

20

TIME

Pressure Control - Synchronized Intermittent Mandatory

Ventilation (PC-SIMV) Mode

The Pressure Control – Synchronized Intermittent Mandatory Ventilation (PC-SIMV) mode provides Spontaneous, Assist, and Mandatory breaths. This mode uses a time window to decide what type of breaths should be delivered. This time window is the maximum time between breaths according to the set Breath Rate.

When you enter PC-SIMV mode, the time window is started. If the patient does not trigger a breath, the ventilator will provide a Mandatory breath when the time expires and then begin the process again. Mandatory breaths deliver the Pressure setting during inhalation and a lower Peak End Expiratory Pressure (PEEP)

during exhalation.