5
Operating | Do not use this device if the ambient temperature is warmer than |
and Storage | 40˚ C (104˚ F). If the device is used at room temperatures warmer than |
Temperatures | 40˚ C, the temperature of the airflow may exceed 43˚ C. This could |
| cause system alarms, thermal irritation, or injury to the patient’s airway. |
Bacteria Filter | Philips Respironics recommends that a main line outlet bacteria filter |
| (Part Number 342077) be used whenever the device is used for invasive |
| therapy or if the ventilator may be used on multiple patients. |
Patient Circuits | The ventilator should only be used with patient interfaces (e.g., masks, |
(General) | circuits and exhalation ports) recommended by Philips Respironics. |
| Proper operation of the device, including alarms, with other circuits has |
| not been verified by Philips Respironics and is the responsibility of the |
| health care professional or respiratory therapist. |
| When adding any components to the breathing system, the flow |
| resistance and dead space of the added components such as |
| humidifiers, speaking valves, Heat Moisture Exchangers (HMEs) and |
| filters should be carefully considered in relation to the potential for |
| adverse effects on the patient’s ventilatory management and device |
| alarms. |
Passive Circuits | An exhalation port is required when using a passive circuit. |
| For the passive circuit, at low expiratory pressures, the flow through |
| the exhalation port may be inadequate to clear all exhaled gas from |
| the tubing – some rebreathing may occur. Rebreathing of exhaled air |
| for longer than several minutes can in some circumstances lead to |
| suffocation. |
Active Circuits | Only use the active exhalation devices designed for Trilogy200. Philips |
| Respironics has not verified proper operation of other active exhalation |
| devices, and their use may result in improper or unsafe device |
| operation. |
| With active exhalation circuits, the exhalation device must be operating |
| properly for the ventilator to deliver therapy. The exhalation device |
| should be inspected on a daily basis and replaced whenever necessary. |
System Checkout | Do not use the ventilator on a patient until a system checkout has been |
| performed. See Chapter 10 of this manual. |
| To make sure the device is operating properly at |
| that the audible tone sounds and the alarm LEDs light red and then |
| yellow momentarily. Contact Philips Respironics or an authorized |
| service center for service if these indications do not occur at |
Chapter 1 Introduction