5

Operating

Do not use this device if the ambient temperature is warmer than

and Storage

40˚ C (104˚ F). If the device is used at room temperatures warmer than

Temperatures

40˚ C, the temperature of the airflow may exceed 43˚ C. This could

 

cause system alarms, thermal irritation, or injury to the patient’s airway.

Bacteria Filter

Philips Respironics recommends that a main line outlet bacteria filter

 

(Part Number 342077) be used whenever the device is used for invasive

 

therapy or if the ventilator may be used on multiple patients.

Patient Circuits

The ventilator should only be used with patient interfaces (e.g., masks,

(General)

circuits and exhalation ports) recommended by Philips Respironics.

 

Proper operation of the device, including alarms, with other circuits has

 

not been verified by Philips Respironics and is the responsibility of the

 

health care professional or respiratory therapist.

 

When adding any components to the breathing system, the flow

 

resistance and dead space of the added components such as

 

humidifiers, speaking valves, Heat Moisture Exchangers (HMEs) and

 

filters should be carefully considered in relation to the potential for

 

adverse effects on the patient’s ventilatory management and device

 

alarms.

Passive Circuits

An exhalation port is required when using a passive circuit.

 

For the passive circuit, at low expiratory pressures, the flow through

 

the exhalation port may be inadequate to clear all exhaled gas from

 

the tubing – some rebreathing may occur. Rebreathing of exhaled air

 

for longer than several minutes can in some circumstances lead to

 

suffocation.

Active Circuits

Only use the active exhalation devices designed for Trilogy200. Philips

 

Respironics has not verified proper operation of other active exhalation

 

devices, and their use may result in improper or unsafe device

 

operation.

 

With active exhalation circuits, the exhalation device must be operating

 

properly for the ventilator to deliver therapy. The exhalation device

 

should be inspected on a daily basis and replaced whenever necessary.

System Checkout

Do not use the ventilator on a patient until a system checkout has been

 

performed. See Chapter 10 of this manual.

 

To make sure the device is operating properly at start-up, always verify

 

that the audible tone sounds and the alarm LEDs light red and then

 

yellow momentarily. Contact Philips Respironics or an authorized

 

service center for service if these indications do not occur at start-up.

Chapter 1 Introduction

Page 16
Image 16
Philips Trilogy 200 Operating, Storage, Temperatures, Bacteria Filter, Patient Circuits, General, Passive Circuits