DC Power Indicator

 

Description

Red Low Battery

When the device detects that an in-use battery’s charge is nearly

Indicator

depleted (has approximately 10 minutes of charge left), the inside

 

of the box surrounding the battery symbol turns red. In addition to

 

the battery indicator on the Monitoring screen, a high priority alarm

 

message will display indicating “Low Battery.” See Chapter 6 for more

 

information. The red indicator is for the last available battery source.

Yellow Battery

Whenever AC power is applied to the device, the internal and

Recharging Symbol ( )

detachable batteries will recharge as needed. If the internal battery is

 

being recharged, the

symbol displays. If the detachable battery is

 

being recharged, the

symbol displays.

Battery Disposal

Dispose of the batteries in accordance with local regulations.

First Time Use

When setting up the device for the first time or after annual service, apply AC power to the ventilator before turning on the blower. Attempting to use the ventilator without first applying AC power, such as installing a detachable battery pack and starting the blower, will cause the internal battery to be displayed in red as an empty battery. When in this state, the internal battery will not be usable until AC power is applied.