38

A sudden change in patient flow will cross the shape signal, causing the pressure level to change.

 

IPAP

 

Pressure

 

 

 

EPAP

 

 

 

Cycle to

 

Shape

EPAP

 

Crossover

 

Signal

 

Point

 

 

Flow

Estimated

 

 

Patient

 

Flow

Trigger to

 

IPAP

 

Crossover

Shape Signal

Point

 

Tracking the patient’s flow pattern with the Shape Signal provides a sensitive mechanism to trigger to inspiration or cycle to expiration in response to changing breathing patterns and circuit leaks.

Trilogy200 clinical manual

Page 49
Image 49
Philips Trilogy 200 manual Shape Signal