Spontaneous/Timed (S/T) Mode

In Spontaneous/Timed (S/T) mode, the device delivers bi-level pressure support. This mode provides Spontaneous and Mandatory breaths. A Mandatory breath is delivered if the patient does not spontaneously breathe within the prescribed Breath Rate (BPM) setting. This ensures that the patient receives a minimum number of breaths per minute. In this mode, an IPAP is delivered during inhalation and a lower EPAP is delivered during exhalation. The duration of a Spontaneous breath is determined by the patient effort. The duration of a Mandatory breath is determined by the inspiratory time setting. The following illustration describes these concepts.

 

 

IPAP = 26 cm H2O

 

1

= Pressure Supported

 

 

 

EPAP = 6 cm H2O

 

 

Spontaneous Breath

 

 

 

Rate = 10 BPM

 

2

= Mandatory Breath

 

 

 

PS = IPAP - EPAP = 20 cm H2O

 

(note longer inspiratory

 

 

 

 

 

time)

 

PRESSURE

 

 

 

 

 

 

 

6 seconds

6 seconds

6 seconds

 

 

 

 

 

 

26

 

 

 

 

 

 

20

 

 

 

 

 

 

 

 

 

 

 

 

S/T Mode

10

1

1

 

2

1

 

6

 

 

 

 

 

 

 

 

 

Enter S/T mode

 

 

 

 

 

 

5

10

15

 

20

TIME

Chapter 3 Modes, Features, and Alarms

Page 34
Image 34
Philips Trilogy 200 manual Spontaneous/Timed S/T Mode