Trilogy200 clinical manual
154
Measured Patient Parameters
Parameter Range Accuracy
Vte/Vti 0 to 2000 ml Greater of 15ml or 15% of reading
Minute Ventilation 0 to 99 l/min Calculation based on measured Vte or Vti
and Respiratory Rate
Estimated Leak Rate 0 to 200 l/min N/A
Respiratory Rate 0 to 80 BPM Greater of 1 BPM or 10% of reading
Peak Inspiratory Flow 0 to 200 l/min 3 l/min plus 15% of reading
Peak Inspiratory Pressure 0 to 99 cm H2O Greater than 2 cm H2O or 10% of reading
Mean Airway Pressure 0 to 99 cm H2O Greater than 2 cm H2O or 10% of reading
% Patient Triggered Breaths 0 to 100% N/A
I:E Ratio 9.9- 1: 1-9.9 Calculation based on Inspiratory Time and
Expiratory Time
All ows and volumes are measured at BTPS conditions.
Spontaneous Breathing During Power Failure Conditions
Flow Set
Point (l/min)
Inspiratory Resistance (cm H2O) Expiratory Resistance (cm H2O)
Active
Circuit
Active Circuit with
Proximal Flow
Sensor
Passive
Circuit
Active
Circuit
Active Circuit with
Proximal Flow
Sensor
Passive
Circuit
30 < 3.0 < 3.5 < 2.0 < 2.0 < 2.0 < 1.5
60 < 9.0 < 10.5 < 5.0 < 3.0 < 4.5 < 4.0
Disposal
Dispose of this device and batteries in accordance with local regulations.