154

Measured Patient Parameters

Parameter

Range

Accuracy

Vte/Vti

0 to 2000 ml

Greater of 15ml or 15% of reading

Minute Ventilation

0 to 99 l/min

Calculation based on measured Vte or Vti

 

 

and Respiratory Rate

Estimated Leak Rate

0 to 200 l/min

N/A

Respiratory Rate

0 to 80 BPM

Greater of 1 BPM or 10% of reading

Peak Inspiratory Flow

0 to 200 l/min

3 l/min plus 15% of reading

Peak Inspiratory Pressure

0 to 99 cm H2O

Greater than 2 cm H2O or 10% of reading

Mean Airway Pressure

0 to 99 cm H2O

Greater than 2 cm H2O or 10% of reading

% Patient Triggered Breaths

0 to 100%

N/A

I:E Ratio

9.9- 1: 1-9.9

Calculation based on Inspiratory Time and

 

 

Expiratory Time

All flows and volumes are measured at BTPS conditions.

Spontaneous Breathing During Power Failure Conditions

Flow Set

Inspiratory Resistance (cm H2O)

Expiratory Resistance (cm H2O)

Point (l/min)

Active

Active Circuit with

Passive

Active

Active Circuit with

Passive

 

Circuit

Proximal Flow

Circuit

Circuit

Proximal Flow

Circuit

 

 

Sensor

 

 

Sensor

 

 

 

 

 

 

 

 

30

< 3.0

< 3.5

< 2.0

< 2.0

< 2.0

< 1.5

 

 

 

 

 

 

 

60

< 9.0

< 10.5

< 5.0

< 3.0

< 4.5

< 4.0

 

 

 

 

 

 

 

Disposal

Dispose of this device and batteries in accordance with local regulations.