Measured Patient Parameters
Parameter | Range | Accuracy |
Vte/Vti | 0 to 2000 ml | Greater of 15ml or 15% of reading |
Minute Ventilation | 0 to 99 l/min | Calculation based on measured Vte or Vti |
|
| and Respiratory Rate |
Estimated Leak Rate | 0 to 200 l/min | N/A |
Respiratory Rate | 0 to 80 BPM | Greater of 1 BPM or 10% of reading |
Peak Inspiratory Flow | 0 to 200 l/min | 3 l/min plus 15% of reading |
Peak Inspiratory Pressure | 0 to 99 cm H2O | Greater than 2 cm H2O or 10% of reading |
Mean Airway Pressure | 0 to 99 cm H2O | Greater than 2 cm H2O or 10% of reading |
% Patient Triggered Breaths | 0 to 100% | N/A |
I:E Ratio | 9.9- 1: | Calculation based on Inspiratory Time and |
|
| Expiratory Time |
All flows and volumes are measured at BTPS conditions.
Spontaneous Breathing During Power Failure Conditions
Flow Set | Inspiratory Resistance (cm H2O) | Expiratory Resistance (cm H2O) | ||||
Point (l/min) | Active | Active Circuit with | Passive | Active | Active Circuit with | Passive |
| Circuit | Proximal Flow | Circuit | Circuit | Proximal Flow | Circuit |
|
| Sensor |
|
| Sensor |
|
|
|
|
|
|
|
|
30 | < 3.0 | < 3.5 | < 2.0 | < 2.0 | < 2.0 | < 1.5 |
|
|
|
|
|
|
|
60 | < 9.0 | < 10.5 | < 5.0 | < 3.0 | < 4.5 | < 4.0 |
|
|
|
|
|
|
|
Disposal
Dispose of this device and batteries in accordance with local regulations.