D.Verify Reset

Wait 40 seconds and verify the following auto-reset conditions:

The High Priority audible indicator has stopped sounding

The red light on the Alarm Indicator/Audio Pause button has stopped flashing

E.Restore Ventilator Settings

Modify the ventilator settings and change the following value shown in Table 8.

Setting

Value

Low Vte/Low Vti

Off

Table 8 - Restore Ventilator Settings

4. Verify Circuit Disconnect Alarm

This procedure verifies that the Circuit Disconnect alarm is working properly. It assumes that you have attached the test lung, verified ventilator settings, and turned on ventilator power as described in the Initial Setup section.

A.Change Circuit Disconnect Ventilator Setting

Modify the Circuit Disconnect ventilator setting to match the value shown below in Table 9.

Setting

Value

Circuit Disconnect

10 seconds

B.Disconnect Test Lung

Disconnect the test lung from the circuit.

C.Verify the Alarm

Wait approximately 10 seconds and verify the following alarm signals:

The High Priority Audible Indicator sounds

A red light flashes on the Alarm Indicator/Audio Pause button

Table 9 - Ventilator Settings

Note: The Low Inspiratory or Low Expiratory Pressure Alarm may also be detected.