A.Change Ventilator Settings

Modify the ventilator settings and change the following values shown below in Table 13.

Setting

Value

Mode

CV

Tidal Volume

500 ml

Breath Rate

12 BPM

Inspiratory Time

1.0 seconds

Flow Pattern

Ramp

PEEP

4 cm H2O

Sigh

Off

Circuit Disconnect

Off

Low Inspiratory Pressure

40 cm H2O

High Inspiratory Pressure

60 cm H2O

Apnea

Off

All other alarms

Off

B.Verify the Alarm

Wait up to 40 seconds and verify the following alarm signals:

The High Priority audible indicator sounds

A red light flashes on the Alarm Indicator/Audio Pause button

The Low Inspiratory Pressure alarm condition appears on the screen, highlighted in red

C.Modify Ventilator Alarm Settings

Modify the Low Inspiratory Pressure setting to match the one shown below in Table 14.

Setting

Value

Low Inspiratory Pressure

6 cm H2O

147

Table 13 - Ventilator Settings

Table 14 - Modify Low

Inspiratory Pressure Alarm

Setting