113
Alarm | Priority | Audible | Visual (Alarm | Device | User/Clinician Action |
|
|
| Indicator Button | Action |
|
|
|
| and Display) |
|
|
Low Inspiratory | High | • • • | • • | Red flashing button; | Operates | User: Press Alarm Indicator/ |
|
| |||||
Pressure |
|
|
| “Low Inspiratory |
| Audio Pause button to |
|
|
|
| Pressure” message |
| silence alarm. Or, press |
|
|
|
|
|
| Left (Reset) button to reset |
|
|
|
|
|
| alarm. Check patient circuit |
|
|
|
|
|
| for leaks or disconnects. If |
|
|
|
|
|
| using an Active PAP or Active |
|
|
|
|
|
| Flow circuit, check proximal |
|
|
|
|
|
| lines to ensure they are |
|
|
|
|
|
| not pinched or have water |
|
|
|
|
|
| condensation (refer to the |
|
|
|
|
|
| Troubleshooting section |
|
|
|
|
|
| – tubing condensation). If |
|
|
|
|
|
| alarm continues, contact your |
|
|
|
|
|
| health care professional. |
|
|
|
|
|
| Clinician: Verify patient |
|
|
|
|
|
| status. If problem continues, |
|
|
|
|
|
| have device serviced. |
|
|
|
|
|
|
|
High Minute | High | • • • | • • | Red flashing button; | Operates | User: Press Alarm Indicator/ |
|
| |||||
Ventilation |
|
|
| “High Minute |
| Audio Pause button to |
|
|
|
| Ventilation” message |
| silence alarm. Or, press Left |
|
|
|
|
|
| (Reset) button to reset alarm. |
|
|
|
|
|
| Continue using device. If |
|
|
|
|
|
| alarm continues, contact your |
|
|
|
|
|
| health care professional. |
|
|
|
|
|
| Clinician: Verify patient |
|
|
|
|
|
| status. |
|
|
|
|
|
|
|
Low Minute | High | • • • | • • | Red flashing button; | Operates | User: Press Alarm Indicator/ |
|
| |||||
Ventilation |
|
|
| “Low Minute |
| Audio Pause button to |
|
|
|
| Ventilation” message |
| silence alarm. Or, press Left |
|
|
|
|
|
| (Reset) button to reset alarm. |
|
|
|
|
|
| Continue using device. If |
|
|
|
|
|
| alarm continues, contact your |
|
|
|
|
|
| health care professional. |
|
|
|
|
|
| Clinician: Verify patient |
|
|
|
|
|
| status. |
Chapter 6 Ventilator Alarms