93
Message | Cause | Action |
Prescription Change Failed | Displayed when an error | Remove the card and have the |
| occurs during the prescription | prescription replaced with a |
| update. This is caused by an | valid prescription. |
| attempt to set a value for a |
|
| prescription parameter that |
|
| is not valid for the therapy |
|
| mode or an attempt to set a |
|
| prescription parameter to an |
|
| invalid value. |
|
Prescription Failed – Circuit | Displayed when the circuit | If the circuit type in the |
Type | type in the new prescription | prescription is correct, remove |
| doesn’t match the circuit type | the card, change the circuit |
| set for the ventilator. | type on the ventilator, and |
|
| |
|
| circuit type on the ventilator is |
|
| correct, remove the card and |
|
| replace the prescription on |
|
| the card with a prescription |
|
| containing the correct circuit |
|
| type. |
Prescription Failed – Card is | Displayed when the | Remove the card and check |
Read Only | prescription is | that the small switch on the |
|
| side of the SD Card is not in the |
|
| Locked position. |
|
| If you continue to receive this |
|
| message, remove the card and |
|
| have the prescription replaced |
|
| with a valid prescription. |
Prescription Failed – Serial | Displayed when the device | Remove the card and have the |
Number | serial number on the new | prescription replaced with the |
| prescription does not match | prescription with the correct |
| the serial number for the | serial number. |
| device. |
|
Prescription Failed – Version | Displayed when the version | Remove the card and have |
| of the prescription does not | the prescription replaced with |
| match the version accepted by | a prescription in the correct |
| the device. | version. |