Prescription Change Failed

Displayed when an error

Remove the card and have the

 

occurs during the prescription

prescription replaced with a

 

update. This is caused by an

valid prescription.

 

attempt to set a value for a

 

 

prescription parameter that

 

 

is not valid for the therapy

 

 

mode or an attempt to set a

 

 

prescription parameter to an

 

 

invalid value.

 

Prescription Failed – Circuit

Displayed when the circuit

If the circuit type in the

Type

type in the new prescription

prescription is correct, remove

 

doesn’t match the circuit type

the card, change the circuit

 

set for the ventilator.

type on the ventilator, and

 

 

re-insert the card. If the

 

 

circuit type on the ventilator is

 

 

correct, remove the card and

 

 

replace the prescription on

 

 

the card with a prescription

 

 

containing the correct circuit

 

 

type.

Prescription Failed – Card is

Displayed when the

Remove the card and check

Read Only

prescription is write-protected.

that the small switch on the

 

 

side of the SD Card is not in the

 

 

Locked position.

 

 

If you continue to receive this

 

 

message, remove the card and

 

 

have the prescription replaced

 

 

with a valid prescription.

Prescription Failed – Serial

Displayed when the device

Remove the card and have the

Number

serial number on the new

prescription replaced with the

 

prescription does not match

prescription with the correct

 

the serial number for the

serial number.

 

device.

 

Prescription Failed – Version

Displayed when the version

Remove the card and have

 

of the prescription does not

the prescription replaced with

 

match the version accepted by

a prescription in the correct

 

the device.

version.