The Circuit Disconnect alarm condition appears on the screen, highlighted in red

D.Reconnect Test Lung

Reconnect the test lung to the circuit.

E.Verify Reset

Wait at least 40 seconds and verify the following auto-reset conditions:

The High Priority audible indicator has stopped sounding

The red light on the Alarm Indicator/Audio Pause button has stopped flashing

F.Restore Ventilator Settings

Modify the ventilator settings and change the following values shown below in Table 10.

Setting

Value

Circuit Disconnect

Off

5. Verify the High Inspiratory Pressure Alarm

This procedure verifies that the High Inspiratory Pressure alarm is working properly. It assumes that you have attached the test lung, verified ventilator settings, and turned on ventilator power as described in the Initial Setup section.

A.Change Ventilator Settings

Modify the ventilator settings and change the following values shown below in Table 11.

Setting

Value

Mode

CV

Tidal Volume

500 ml

Breath Rate

12 BPM

Inspiratory Time

1.0 seconds

Flow Pattern

Ramp

145

Table 10 - Restore Ventilator Settings

Table 11 - Ventilator Settings

Note: If this alarm is not reset within 3 occurrences, the alarm is elevated to High Priority, and the High Priority Indicators occur.

Chapter 10 System Checkout Procedures

Page 156
Image 156
Philips Trilogy 200 manual Verify the High Inspiratory Pressure Alarm, Setting Value Circuit Disconnect Off, 145