67

Indicator

Description

Low Pressure

If you enable a volume therapy mode, this indicator appears

Indicator

L

below the manometer bar, indicating the low pressure alarm

 

 

 

 

setting.

High Pressure

If you enable a volume therapy mode, this indicator appears

Indicator

H

below the manometer bar, indicating the high pressure alarm

 

 

 

 

setting.

Pressure

 

 

 

This indicator displays the current patient pressure. This only

 

 

 

 

appears in detailed view.

Respiratory Rate (RR)

This indicator displays the measured respiratory rate in Breaths

 

 

 

 

Per Minute (BPM). This only appears in detailed view.

Exhaled Tidal

This indicator displays the estimated exhaled tidal volume in

Volume (Vte)

milliliters and reflects compensation for BTPS. This only appears in

 

 

 

 

detailed view when Passive Circuit is selected.

Inhaled Tidal

This indicator displays delivered tidal volume in milliliters and

Volume (Vti)

reflects compensation for BTPS. This only appears in detailed view

 

 

 

 

when the Active with PAP Circuit is selected.

Leak

 

 

 

This indicator displays the total leak (non-returned flow), between

 

 

 

 

the unit outlet and the patient, averaged over the previous

 

 

 

 

breath. This only appears in detailed view when the Passive

 

 

 

 

Circuit is selected.

Chapter 5 Viewing and Changing Settings

Page 78
Image 78
Philips Trilogy 200 manual Per Minute BPM. This only appears in detailed view