Alarm | Priority | Audible | Visual (Alarm | Device | User/Clinician Action |
|
|
| Indicator Button | Action |
|
|
|
| and Display) |
|
|
Keypad Stuck | Low | • • | Solid yellow button; | Operates | User: Press Alarm Indicator/ |
| |||||
|
|
| “Keypad Stuck” |
| Audio Pause button to |
|
|
| message. |
| silence alarm. Or, press Left |
|
|
|
|
| (Reset) button to reset alarm. |
|
|
|
|
| Check the keys to determine |
|
|
|
|
| if they are lodged in the case. |
|
|
|
|
| If alarm continues, place |
|
|
|
|
| patient on alternate source of |
|
|
|
|
| ventilation and contact your |
|
|
|
|
| health care professional. |
|
|
|
|
| Clinician: Check keys |
|
|
|
|
| to determine if they are |
|
|
|
|
| lodged in the case. If alarm |
|
|
|
|
| continues, place patient |
|
|
|
|
| on alternate source of |
|
|
|
|
| ventilation and have device |
|
|
|
|
| serviced. |
|
|
|
|
|
|
Battery | Info | • | “Batt Discharge | Operates | User: Press Alarm Indicator/ |
| |||||
Discharging |
|
| Stopped – Temp.” |
| Audio Pause button to |
Stopped Due to |
|
| message |
| silence alarm. Or, press |
Temperature |
|
|
|
| Left (Reset) button to reset |
|
|
|
| alarm. Move device to a | |
|
|
|
|
| |
|
|
|
|
| cooler location. Make sure |
|
|
|
|
| device is not close to a heat |
|
|
|
|
| source. Make sure cooling |
|
|
|
|
| vents are not blocked. If |
|
|
|
|
| alarm is detected for internal |
|
|
|
|
| battery and continues, place |
|
|
|
|
| patient on alternate source of |
|
|
|
|
| ventilation and contact your |
|
|
|
|
| health care professional. |
|
|
|
|
| Clinician: If alarm is detected |
|
|
|
|
| on the internal battery and |
|
|
|
|
| continues after the measures |
|
|
|
|
| above have been taken, place |
|
|
|
|
| patient on alternate source of |
|
|
|
|
| ventilation and have device |
|
|
|
|
| serviced. |