Chapter 3 Modes, Features, and Alarms

21
Therapy Mode Table
The following table summarizes all of the therapy modes and the settings available in each mode. Some settings in the table
are dependent upon other settings. For instance, if the circuit type is set to Active with Flow, then the Flow Trigger Sensitivity,
Leak Compensation, and Flow Cycle settings will display.
Note: Pressure Support, referred to in the table below and later in this manual, is dened as IPAP - EPAP or Pressure -
PEEP (PC-SIMV).
Therapy Modes
CPAP S S/T T PC PC-SIMV CV AC SIMV
Dual Prescription

Therapy Parameters

Circuit Type √ √ √ √ √ √
CPAP √
IPAP √ √
AVAPS (On, O)*
IPAP Max Pressure
IPAP Min Pressure
EPAP √ √ √ √
Pressure √
Pressure Support (PS)
PEEP √ √
Tidal Volume √ √ √ √ √
Breath Rate
Inspiratory Time √ √ √ √ √
Trigger Type* √ √ √ √ √
Flow Trigger Sensitivity √ √
Leak Compensation
Flow Cycle Sensitivity
Ramp Length
Ramp Start Pressure
Flex *
Rise Time √ √
Flow Pattern
Sigh √ √ √
Circuit Disconnect
Apnea √ √ √ √ √
Apnea Rate
High Vte* √ √ √ √ √ √ √
Low Vte* √ √ √
High Vti* √ √ √ √ √
Low Vti* √ √
High Minute Ventilation √ √ √ √ √ √ √
Low Minute Ventilation √ √
High Respiratory Rate
Low Respiratory Rate
High Inspiratory Pressure
Low Inspiratory Pressure
* Trigger Type, AVAPS, and Flex settings are only available with the Passive circuit type. High Vte and Low Vte settings are only
available for Passive and Active Flow circuit types. High Vti and Low Vti settings are only available with the Active PAP circuit
type. Leak compensation is only available with the Active Flow circuit type.