27

Volume Control Ventilation Therapy Modes

Volume Control ventilation modes deliver a prescribed inspired tidal volume according to a set Breath Rate and a set Inspiratory Time.

Control Ventilation (CV) Mode

In Control Ventilation (CV) mode, the device delivers volume control therapy. This mode provides only Mandatory breaths. A Mandatory breath is delivered according to the prescribed BPM setting for the prescribed Inspiratory Time. This also means that the ventilator will not respond to patient effort. In this mode, the Tidal Volume is delivered during inhalation and PEEP is delivered during exhalation. The following illustration describes these concepts.

FLOW

50

0

-50

 

Tidal Volume = 500 mL

2

= Mandatory Breath

 

 

Rate = 10 BPM

 

 

 

 

 

Inspiratory Time = 1.5 seconds

 

 

 

Flow Pattern = Square

 

 

 

6 seconds

6 seconds

 

6 seconds

6 seconds

 

Enter CV mode

 

 

 

 

Control Mode

2

10

2

15

2

 

5

 

 

 

20

TIME

Chapter 3 Modes, Features, and Alarms

Page 38
Image 38
Philips Trilogy 200 manual Volume Control Ventilation Therapy Modes, Control Ventilation CV Mode